Nicely explained. Thanks a lot. But what if the electrode has less efficiency i.e. slope range is less than 90%. How can we increase the slope range from let's say 85% to 98%?
Slope issues & how to overcome it The % obtained can tell you if the slope is good. It should between 92-102%. If the slope is not within expectation, User may recheck again as below: • Temperature probe : Is the customer using a temperature probe. Because temperature can influence the pH values. (See attachment) • Buffer issue: Identify which buffer mV value is not within expectation. Change a different lot/batch • Electrode contamination: Clean the electrode with the appropriate cleaning method. Flush and refill a new batch of fill solution in the electrode.
Slope issues & how to overcome it The % obtained can tell you if the slope is good. It should between 92-102%. If the slope is not within expectation, User may recheck again as below: • Temperature probe : Is the customer using a temperature probe. Because temperature can influence the pH values. (See attachment) • Buffer issue: Identify which buffer mV value is not within expectation. Change a different lot /batch • Electrode contamination: Clean the electrode with the appropriate cleaning method. Flush and refill a new batch of fill solution in the electrode.
Good video!! 👍🏻👍🏻 It helps a lot
Very helpful!
Very Useful video..
Informative 👍
Very Very easily Calibration
Thanks 😊 🙏 😊
Thank you for explaining. In our case calibration point not detected. How to resolve the issue
Well done
I do not have an ATC probe. Will it work for pH tests?
what to do if PH electrode is turned brownish inside
Nicely explained. Thanks a lot.
But what if the electrode has less efficiency i.e. slope range is less than 90%. How can we increase the slope range from let's say 85% to 98%?
Slope issues & how to overcome it
The % obtained can tell you if the slope is good. It should between 92-102%.
If the slope is not within expectation, User may recheck again as below:
• Temperature probe : Is the customer using a temperature probe. Because temperature can influence the pH values. (See attachment)
• Buffer issue: Identify which buffer mV value is not within expectation. Change a different lot/batch
• Electrode contamination: Clean the electrode with the appropriate cleaning method. Flush and refill a new batch of fill solution in the electrode.
My dear sir
Please make one video on BTG consistency transmitter.
Thanks
If error occurre due to slp then
How to change ph 9 insted of 10
Please provide the catalogue number for this meter.
And the pH electrode order number.
I am sure you are a female because your hand is so beautiful like a girl
What about slope
Slope issues & how to overcome it
The % obtained can tell you if the slope is good. It should between 92-102%.
If the slope is not within expectation,
User may recheck again as below:
• Temperature probe : Is the customer using a temperature probe. Because temperature can influence the pH values. (See attachment)
• Buffer issue: Identify which buffer mV value is not within expectation. Change a different lot
/batch
• Electrode contamination: Clean the electrode with the appropriate cleaning method. Flush and refill a new batch of fill solution in the electrode.
@@andylownav285 k thank you sir, when slope is below 92% is calibration is passed or not
And also tell buffers acceptable ratio