How to Calibrate and use the pH Meter?

Поділитися
Вставка

КОМЕНТАРІ • 89

  • @d.deepak
    @d.deepak 4 роки тому +6

    great explanation, got my pH meter calibrated with your help. Thanks

  • @chemfunman
    @chemfunman  4 роки тому +2

    SUBSCRIBE NOW to this channel now to receive more great chem videos!
    Connect with me on
    Facebook | facebook.com/chemfunman/
    Twitter @chemfunman (twitter.com/chemfunman)
    Stay safe!
    “Ever tried. Ever failed. No Matter. Try again. Fail again. Fail better.”
    - Samuel Beckett

  • @guncix21
    @guncix21 3 роки тому +1

    When you take the KCl conditioning vial of the pH probe, you have to hold the clear vial part and twist the white part anti-clockwise, therefore the vile will be gently released. When you finish analysis put it back on and twist it clockwise direction to seal it.

  • @afzalshahbaz89
    @afzalshahbaz89 5 років тому +5

    at 2:05 you don't have to pull out the probe from the 3M KCL soaking tube. Simply rotate the upper flexible screw of the soaking tube and the probe will come out itself without any forced rotation of the soaking tube from the bottom. Else it may damage the probe bulb in process.

    • @chemfunman
      @chemfunman  4 роки тому

      afzal shahbaz that’s right!

  • @jomarcamu7057
    @jomarcamu7057 5 років тому +5

    Seriously do you think you did it right? You calibrated your instrument by standard buffer 4.01, followed by 9.21 and finally by 7.00. And after calibration you pressed exit button without even pressing END button hence not saving your calibration into your instrument. Consequently no slope can be displayed by your instrument since you didn't even ended your calibration and saved it. Correct me if I am wrong if you disagree.

    • @hafizusamma1497
      @hafizusamma1497 5 років тому

      Good Question. Would you please, refer me some of best companies to purchase a Ph meter? One reliable, robust too for industrial use. Thanks! (hafizusamma@mpcl.com.pk)

    • @ihartu5549
      @ihartu5549 Рік тому

      he calibrated the buffer ph7 first 😏 lmao don't comment when you can't comprehend.. and one more thing, not all ph meter are the same...

  • @jhogan1960
    @jhogan1960 5 років тому +3

    Interesti9ng how they do things in different countries. I always learn something New. Hello from the USA!

    • @chemfunman
      @chemfunman  5 років тому

      appreciate your comment! hello from Singapore, James!

  • @pavadaiarumugam9723
    @pavadaiarumugam9723 3 роки тому +2

    Very clear explanation. Thank you so much sir

  • @ashrafgafar
    @ashrafgafar 3 роки тому +2

    This was brilliant

    • @chemfunman
      @chemfunman  3 роки тому +1

      Thank you so much for your support. God bless and subscribe, share this love or learning with others. Cheers.

  • @mightyman6459
    @mightyman6459 6 років тому +3

    Very useful video, thanks teacher!

  • @medosou
    @medosou 5 років тому +2

    Very well demonstrated

  • @xavik
    @xavik 8 років тому +1

    Thanks! I already missed these videos.

    • @chemfunman
      @chemfunman  8 років тому

      you are most welcome :')

  • @JashanjotKaur
    @JashanjotKaur 3 роки тому +1

    Very well explained. Thanks.

    • @chemfunman
      @chemfunman  3 роки тому +1

      thank you very much! please subscribe and share this video with your friends!

  • @thembelangxokile1571
    @thembelangxokile1571 7 років тому +3

    it would be helpful though if you told us why we do certain things like taking steps to remove the bubbles

    • @chemfunman
      @chemfunman  7 років тому +1

      would you like to chip in , appreciate if all do out part ! ^^

  • @Sp-fe1eo
    @Sp-fe1eo 2 роки тому +1

    thanks for the video...it's really useful

  • @MaestroAssistant
    @MaestroAssistant 4 роки тому +1

    I think you can rotate white plastic top to unscrew it from the KCl container

    • @chemfunman
      @chemfunman  4 роки тому

      Daniel Arroyo you are right!

  • @matchmedicines8209
    @matchmedicines8209 Рік тому +1

    thanks a lot, it was amazing.
    just a question, how I can prepare the buffers (4,9, and 7)?

    • @chemfunman
      @chemfunman  Рік тому

      we bought them from the store.

  • @amenananias2134
    @amenananias2134 Рік тому +1

    A good video

  • @yasiralshebib6626
    @yasiralshebib6626 3 роки тому +2

    Hi Dr Fun
    thanks for the education video.
    I'd like to ask you about: (#good safety habit), is it just a hashtag on twitter or there is a youtube video channel dedicated to safety habits videos?
    thanks in advance

  • @yohanseol6296
    @yohanseol6296 6 років тому +1

    thank you very much!!^^

  • @lydiaamelia199
    @lydiaamelia199 3 роки тому +2

    Great and clear explanation, i love that, but I have a question, what we have to do if the pH meter measure wrongly or inaccurate when calibrating the buffer solution, like when the actual pH value of the buffer is 4.01 but the pH meter show the value of 4.51 then what should we do?

    • @chemfunman
      @chemfunman  3 роки тому

      plot a calibration curve, and adjust accordingly

  • @parimisantu5148
    @parimisantu5148 7 років тому +1

    Nice

  • @alial-absawi8880
    @alial-absawi8880 6 місяців тому +1

    Hello
    I hope you are doing well
    May I ask that why u didn't press End after that Save
    Also I have problem with my ph meter which one of the puffer always can not be celebrated

    • @chemfunman
      @chemfunman  6 місяців тому +1

      I don’t understand your last sentence

    • @alial-absawi8880
      @alial-absawi8880 6 місяців тому

      Sorry I meant calibrated @4&7ok but when I come to 10 display showed out of the range

  • @ZeMagnifier
    @ZeMagnifier 9 місяців тому +1

    Won't we risk contaminating the buffer solutions if we were to directly dip the proble into the buffer solution storage?

    • @chemfunman
      @chemfunman  9 місяців тому

      can wipe quickly

    • @ZeMagnifier
      @ZeMagnifier 9 місяців тому

      @@chemfunman I don't mean the probe itself, I mean the buffer solution. Is it still as accurate as the original measurements before callibrating the meter?

  • @longho2097
    @longho2097 4 роки тому +1

    when i select Menu structure of pH/ion, the display only show two mode 1 Sensor ID/ IM and 2 Calibration Setting. I choose mode 2 but the machine don't show mode Buffer Group/ Standard. On the other hand, at the main display, the machine in Router mode, i can't calibration the group buffer 4, 7, 9.21 because this machine mesure not exactly. Can you help me the method to this machine return as originally, when i press menu structure of pH/ion have 7 mode. Thank you!!

  • @wafakallaa5152
    @wafakallaa5152 7 років тому +1

    thanks a lot

  • @alyaae.h3533
    @alyaae.h3533 7 років тому +1

    thank you...

  • @shanrao5388
    @shanrao5388 3 роки тому +1

    What type of error gave the PH meter to find out PH..?

  • @bouallalahalima6459
    @bouallalahalima6459 2 роки тому +1

    Thank's but I have a problem when l calibration read only one calibration 1 don't read calibration 2

  • @muntahafazal237
    @muntahafazal237 5 років тому +1

    Wow..

    • @chemfunman
      @chemfunman  5 років тому

      Muntaha Fazal that’s an old game

    • @chemfunman
      @chemfunman  5 років тому

      Muntaha Fazal legendary

  • @RajendraSingh-ps6of
    @RajendraSingh-ps6of 5 років тому +1

    How to plot pH graph with different buffer solution

  • @saikorada8342
    @saikorada8342 11 місяців тому +1

    Sir telugu mediam sir please

  • @uartim
    @uartim Рік тому +1

    Do you need a buffer with higher pH value for solution to be measured at a higher pH ?

  • @jaswanthchowdary17-9-21
    @jaswanthchowdary17-9-21 6 років тому +1

    Sir I want slope calculation on 3 buffers calibration ND u miss the slope value in this video...

    • @chemfunman
      @chemfunman  3 роки тому

      slope is the measure of the linearity of the calibration - is it accurate? can we believe the readings? with what type of precisions?

  • @suppachaikrajangsri2196
    @suppachaikrajangsri2196 3 роки тому +1

    Hello, I have a problem with Cal2, it show wrong buffer, what could be wrong?

    • @chemfunman
      @chemfunman  2 роки тому

      the comments show "wrong buffer"? check if the buffer is new. because it is prone to pH change when used for a long time.

  • @cosmeticreviews9
    @cosmeticreviews9 4 роки тому +1

    Please tell what does offset and slope mean in pH meter?

    • @chemfunman
      @chemfunman  3 роки тому

      offset is the variance - inherent error by any instrument

    • @chemfunman
      @chemfunman  3 роки тому

      slope is the measure of the linearity of the calibration - is it accurate? can we believe the readings? with what type of precisions?

  • @jerilyncruz5069
    @jerilyncruz5069 4 роки тому

    Do you only need to calibrate once or every use? Or just after every session?

  • @SVERMA941
    @SVERMA941 5 років тому +1

    What is purpose of slop and offset limit?

    • @chemfunman
      @chemfunman  3 роки тому +1

      slope is the measure of the linearity of the calibration - is it accurate? can we believe the readings? with what type of precisions?

    • @SVERMA941
      @SVERMA941 3 роки тому +1

      @@chemfunman After calibration u can verify by buffer

  • @hamzaelfadili-7875
    @hamzaelfadili-7875 6 років тому +1

    please what's the role of Kcl

    • @chemfunman
      @chemfunman  6 років тому

      prevent leaching of water via osmosis. if this happens, the KCl crystallize. forming an ionic salt. immobilized ions

    • @chemfunman
      @chemfunman  6 років тому

      prevent leaching of water via osmosis. if this happens, the KCl crystallize. forming an ionic salt. immobilized ions

    • @chemfunman
      @chemfunman  6 років тому

      prevent leaching of water via osmosis. if this happens, the KCl crystallize. forming an ionic salt. immobilized ions

  • @agustinnovalinda6468
    @agustinnovalinda6468 2 роки тому

    Can u explain hau to change temperature setting from fareinheit to celcius on display, please share

  • @mimikiongmeimei6659
    @mimikiongmeimei6659 5 років тому +1

    why use pH 7 first?

    • @chemfunman
      @chemfunman  3 роки тому

      neutrality is always unbiased. that is the best reference.

  • @myassarkhiami6442
    @myassarkhiami6442 3 роки тому +1

    I need this machine

  • @jawadnadeem2920
    @jawadnadeem2920 5 років тому +1

    what if my first two calibrations go well and the third one shows 10.01 instead of 9.21?? what could go wrong?? possibly it does not measure properly above 7???

  • @Joshua-dc1bs
    @Joshua-dc1bs 6 років тому +2

    My favourite part of my job is staring at the screen as it... slowly... calibrates. Weighing scales are even worse :/

  • @shi__532
    @shi__532 4 роки тому +2

    haha Slam dunk enthusiast??

    • @chemfunman
      @chemfunman  4 роки тому

      Syifaq Azlan not bad you recognise!

    • @chemfunman
      @chemfunman  4 роки тому +1

      Syifaq Azlan very good series

    • @shi__532
      @shi__532 4 роки тому +1

      @@chemfunman childhood anime