  • Video explaining standard additions calibration, including how it is performed and how to calculate the concentration of an unknown solution. Presented by Dr Daniel Belton, University Teaching Fellow, University of Huddersfield. Made with support from the Analytical Chemistry Trust Fund.

КОМЕНТАРІ • 22

  • @youdontwannaknow3397
    @youdontwannaknow3397 3 роки тому +10

    You’re a life saver. Thank you. This was so helpful.

  • @iremcakar6545
    @iremcakar6545 2 роки тому +5

    best and cleariest and absolute explanation ever, thank you so much! :D

  • @hilarylee20
    @hilarylee20 3 роки тому +2

    Very helpful, needed this refresher. Much Thanks! :)

  • @Whiteblack171
    @Whiteblack171 3 роки тому +1

    Thank you , this is very helpful !

  • @the-state-of-the-art4967
    @the-state-of-the-art4967 3 роки тому

    This makes sense! good explanation.

  • @randeepkaur9881
    @randeepkaur9881 3 роки тому +1

    Helpful video.
    U tell me meaning of spike....
    thank u very much

  • @brianapomales3775
    @brianapomales3775 2 роки тому

    Thank youuu!!

  • @NGUYENQUANGKHAIINVEST
    @NGUYENQUANGKHAIINVEST 2 роки тому

    thank you!

  • @mijanrahaman2808
    @mijanrahaman2808 Рік тому

    Thank you very much. It is very helpful

  • @lyn4713
    @lyn4713 3 роки тому +6

    Why is it multiplied by 10?

    • @MAngelier
      @MAngelier 2 роки тому +8

      Because each time he added 10mL of the sample in a total volume of 100mL, so it gets 10 times diluted. The measured value is therefore actually 10x to low, to compensate for this, multiply by 10 in the end.

  • @drdeepaksoni6639
    @drdeepaksoni6639 2 роки тому

    i found more area of analyte than standard in egg matrix ( LC-UV analysis)
    sample 5 g
    reconstitute volume 2ml
    spiked conc 1 ppm
    how do i calculate conc of analyte to find out recovery?
    area of sample 57289
    area of standard 42380

  • @igsr1869
    @igsr1869 Рік тому

    Thank you. It was very useful. How can we find the percentage recovery?

    • @ChemEngTutor
      @ChemEngTutor  Рік тому

      To do this, you need to know the concentration of analyte you were expecting to get. You can then work out the experimentally determined concentration as a percentage of the expected concentration. I hope this helps.

  • @mimosaya4462
    @mimosaya4462 Рік тому

    Hi, just wanna ask. Why do we need to multiply 10 to the calculated concentration? (As per the example calculation)

    • @ChemEngTutor
      @ChemEngTutor  Рік тому

      We need to multiply by 10 because the initial specimen solution was diluted by a factor of ten when the standards were made.

  • @LOLA-yj9rr
    @LOLA-yj9rr Рік тому

    Why do we extrapolate it to when the instrument response is zero? Doesn't a zero response mean the absence of analyte?

    • @ChemEngTutor
      @ChemEngTutor  Рік тому

      We are trying to get the difference between when there is zero analyte (y = 0) and when we just at have the instrument response for the specimen (x = 0). I hope this helps.

  • @omerakrawi8729
    @omerakrawi8729 2 роки тому

    from where 0.506 ???

    • @funnysundayfunnys8537
      @funnysundayfunnys8537 2 роки тому

      Use y=mx+c formula, 0.506 in this case is the gradient/slope of the sample.

    • @matthulvey8615
      @matthulvey8615 Рік тому

      All values of the calibration plot are from least squares regression which finds the algebraic formula that best fits the raw data. The constants are merely the slope and intercept with x and y as your variables. Typically you would use software to calculate, but you can do so manually.

    • @ericdominic8382
      @ericdominic8382 Рік тому

      Wondering too