Metrology Quality Rules Tur-Tar

Поділитися
Вставка
  • Опубліковано 30 лис 2024

КОМЕНТАРІ • 10

  • @465maltbie
    @465maltbie 6 років тому +1

    I must have missed something but where did the 10 min number come from? How did you determine that the zero gage block was within that calculation? I listened twice and I think I still must have missed it.

    • @dr.jimsalsbury695
      @dr.jimsalsbury695 6 років тому +7

      The ±10 µin uncertainty was taken from an example in the ASME B89.1.13 standard. How to estimate uncertainty is beyond the scope of this video, so we just borrowed the value from a recognized source. In order to do TUR calculations, you first need uncertainty. To see how the 10 µin uncertainty was determined, please check out ASME B89.1.13 (in the appendix on measurement uncertainty). Also, see www.mitutoyo.com/mitutoyo-on-demand/decision-rules-tar-and-tur/

    • @465maltbie
      @465maltbie 6 років тому +2

      Thank you I heard that in the video but I guess I just didnt understand it. I will check out the links. Charles

  • @tropicalcasa
    @tropicalcasa 6 років тому +7

    Excellent presentation and use of the correct terms of TUR and TAR. I like that presenter was very objective and professional in terms of the application of the calibrated instrument.

    • @dr.jimsalsbury695
      @dr.jimsalsbury695 5 років тому +3

      It is comments like this that motivates us to make more videos. Thank you so much.

  • @davidbeier1562
    @davidbeier1562 4 роки тому +2

    When calculating the TUR for determining a ratio 4 to 1, is the resolution of the TI included (0.6R) or repeatability? A 4 to 1 will never be achieved using that method. I was at a NCSL seminar in Cleveland and the presenters said that the resolution should be included. I recall Jim standing up and say it shouldn't only for your CMC but not when determining a TUR. Certain accredited bodies require the resolution be used to calculate TUR. I have seen companies state a process TUR which doen't include any variables resolution or repeatability . What is it and where is the guidance document to prove it one way or another.

    • @boatmanfall3n
      @boatmanfall3n 2 роки тому +1

      JCGM 100:2008 or otherwise known as "The GUM" (Guide to the expression of uncertainty in measurement)

  • @israelbarrios1549
    @israelbarrios1549 5 років тому +3

    excellent

  • @gajanank3812
    @gajanank3812 4 місяці тому

    Nice

  • @xddrek
    @xddrek 5 років тому

    v.good.
    I couldn't understand that that some accreditation body giving accreditation of a lab have their CMC +-2% in pressure measurement. Then what is the use of that lab. whether ILAC have any guideline on that?