Hi, I have a testo 184 T3 that apparently was working fine but I just took the report and configured it but I press start and it does not start recording the temperatures, it has enough battery and I can see it is recording the ambient temperature but it does not record. what could be what is happening? thanks.
Do we nees to configure it after every use/recording? Because when I press stop, it goes to end. But how to go back to wait for another session of recording? If we need to configure it every time, that is pitty.
@ariloulchillhop hi. After 2 years of use, I must told you that after every recording session you need to put it on PC, delete old configuration and configure again and save it. Don't forget to click on NFC if you're using a Testo printer. Otherwise it will not print. I recommend to use Expert mode for configuring. BR
Thank you for your nice explanation. I do have Testo 184 T3 and would like to export data into Excel format. but I don't have the software which that device can recognize. can you help on to find proper software for it?
You use the Testo 'Comfort Pro' software to create a CSV/Excel file ... Unfortunately this is a purchased item from Testo. (See around 4:45 into video)
You simply run the 'configuration' PDF again to start a new test. This unit does not have a start, stop then start again ability. You can get a 'summary' using the NFC interface without actually stopping the unit. Its a pretty neat box with a pretty neat interface.
Hi Graeme, Thank you. I was trying to configure using an another PDF reader instead of Adobe and hence the issue. I have already performed the reconfiguration with Adobe. Any way, thanks a lot for your time.
Hi, I have a testo 184 T3 that apparently was working fine but I just took the report and configured it but I press start and it does not start recording the temperatures, it has enough battery and I can see it is recording the ambient temperature but it does not record. what could be what is happening? thanks.
Do we nees to configure it after every use/recording? Because when I press stop, it goes to end. But how to go back to wait for another session of recording? If we need to configure it every time, that is pitty.
I really want to know the answer for this
@ariloulchillhop hi. After 2 years of use, I must told you that after every recording session you need to put it on PC, delete old configuration and configure again and save it. Don't forget to click on NFC if you're using a Testo printer. Otherwise it will not print. I recommend to use Expert mode for configuring. BR
Thank you for your nice explanation. I do have Testo 184 T3 and would like to export data into Excel format. but I don't have the software which that device can recognize. can you help on to find proper software for it?
The software we supply for our rental products is Testo ComSoft PRO version 4.3
excuse me, I have 184 H1 and I would like to export data into excell format. would you recommend me a suitable software please.😊
You use the Testo 'Comfort Pro' software to create a CSV/Excel file ... Unfortunately this is a purchased item from Testo. (See around 4:45 into video)
@@GraemeMurphy54 thank you so much
That was a nice tutorial, Thank you.
Can you please guide me, how to start again once you end one test.
You simply run the 'configuration' PDF again to start a new test. This unit does not have a start, stop then start again ability. You can get a 'summary' using the NFC interface without actually stopping the unit. Its a pretty neat box with a pretty neat interface.
Hi Graeme,
Thank you.
I was trying to configure using an another PDF reader instead of Adobe and hence the issue.
I have already performed the reconfiguration with Adobe. Any way, thanks a lot for your time.