If I am using a condensate pot in LP line then what will the calibration procedure?? Because when we are using condensate pot in LP line then liquid present in LP line filled up to condensate pot exerts a constant pressure on transmitter diaphragm so we will not able to use this procedure i.e. venting LP to environment. Hoping for your feedback regarding query.
Use digital hand pump or low range druck, no need to connect the gauge. Also at low range pressure, adjust zero by matching the hart mA reading with multimeter reading,.Do the same for 20 mA with high pressure range.
Sir aap se bahut sara knowledge sikhna hae r bataana aap hi ko hae.ho sake to loop folder documents koun sa kis kaam ke lye aata hae wo batayega please
Brother Can You Adjust zero & span through HART ? What I know HART is a communication protocol, it's not a calibrator. We can adjust zero through HART but not calibrate. I don't know how u r calibrating through HART.
Sir, First, Thanks for this nice and helpful tutorial, also, kindly use English if it is possible, to make your videos useful for every one around the globe. Second, even if you don't use any 1-5 VDC measuring device here, you need to have a loop resistor in order to maintain the communication between the transmitter and the HART communicator. If you bypass the resistor, the power supply filter capacitors will eat your HART signals (AC)..! [By an instrumentation infant and an electronics teen :) ]
Mr. mohammad equaluser vavle will be close here if u will not close equaliser then ur given pressure through hp line is also will also go lp line through equliser nd u find zero pressure in output
Maximum pressure is supposed 10 bar.....and calibration range is 0 to 6 bar.....it means your tx is calibrated 0 to 6 bar....if you want calibrate this tx from 0 to 10 bar you can bcoz maximum pressure is 10 bar it's diaphragm capacity
सर आपको बहुत बहुत धन्यवाद। आपका ये विडियों देखकर इनटरभियू देने गया और पास कर गया और नौकरी मिल गई है। केवल जवानिंग बाकी है।
आप वीडियो बहुत अच्छा बनाते हो सर थैंक यू
Thank u so much to explain dp calibratino in hindi language....sometimes dificult to understand video in english language so once again thanks for this video in hindi
Thanks sir 100 % pure btate hain
आपके video बहुत helfful होते है
all are good
Perfect Explanation
Thanks for knowledge Given simple and easy Hindi language
very nice and informative video sir, thank you very much for easy explanation in video.Best regards, Mohan Singh
Thank You, Mohan Singh ji,
very well done sir... it,s really very nice and useful video thanks for this..
very gud work
Thank you somuch
Wow very nice video sir ,lots of thank u
Apko automation ka bhi itna knowledge hai pata nhi tha ...please or video automation pe hi banayiye 🙏
perfect way to get this thing fit in my mind. thank you👊
If I am using a condensate pot in LP line then what will the calibration procedure??
Because when we are using condensate pot in LP line then liquid present in LP line filled up to condensate pot exerts a constant pressure on transmitter diaphragm so we will not able to use this procedure i.e. venting LP to environment.
Hoping for your feedback regarding query.
Hello sir, can u please tell me why we keep LP side open to atmosphere ?
Very important question aap sir bataye thanks
the explication is very good very important for the work of the instrumentation
Thank You Very Much Nelsom Hermandez Ji
Sir power source ke series 250ohm resistance lagega ya HART ke series me.
Very nice explanation ..
nice..much helpfull.
What about if we don't have to adjust it via Hart but with range of foxbro only..also are der any other videos regarding this topic
Sir please solve my question dpt close tank me urv positive range kyo le raha hai
Excellent
Thank you so much sir
bahot bahot dhanyavad sir....
Thank You, Rajesh Kumar Ji
nice vedio sir........
Thank You, Sandeep Kumar Ji
Awesome sir...
Thank You Very Much Mohammed Asgar Shikh Ji
Use digital hand pump or low range druck, no need to connect the gauge.
Also at low range pressure, adjust zero by matching the hart mA reading with multimeter reading,.Do the same for 20 mA with high pressure range.
Very useful video..Sir if I have no Hart communicator , and no pressure pump.then how I calibrate DPT..Please reply
Nice Explanation bro
Thank You Yasir Sir Ji
Nice video sir
How we will configured the scale variable by hart communicator?? Please guide us, Video is very helpful ....
visit ua-cam.com/channels/d6LP1BncR9AZbPf9-dLNCA.html
Sir aap se bahut sara knowledge sikhna hae r bataana aap hi ko hae.ho sake to loop folder documents koun sa kis kaam ke lye aata hae wo batayega please
Thanks sir
Control valve ka bhi calibration video bnayye sir
Sir Hart communicator video banao sir 💥👍🏼 super fainstatic
Thanks for previous video .Sir trouble shooting me v ek video banaye...please
Sir hum transmitter ko series me kitne display karwa sakte hai
excellent
Dear Superb!
Thanks, Furah Shaikh Ji
Super
Brother Can You Adjust zero & span through HART ? What I know HART is a communication protocol, it's not a calibrator. We can adjust zero through HART but not calibrate. I don't know how u r calibrating through HART.
visit ua-cam.com/channels/d6LP1BncR9AZbPf9-dLNCA.html
Good job
sir kya apne kevl ek hi instument pr video bnaya h hai? agr aur bhi bnaya h to uska link bhejiye sir...
very good
calibration
i think 250ohms hart ko online hone ke liye v to hota hai n sir
Why LP side atmosphere???? Sir plz reply me
FT LT TT ka Bhi Video banaiye
Thank you ❤
Sir one more thing..Please upload video on Level Transmitter , Flow Transmitter , Temperature Transmitter ' s calibration..
OK,Rajnish Singh Ji, We will upload very soon
visit ua-cam.com/channels/d6LP1BncR9AZbPf9-dLNCA.html
Thanks a lot sir
Sir thanks you.very nice
Your Welcome, Sanjeeb ji
zero & span agar kaam nhi kr rha to Kya problem ho sakta hai please give me suggestions
very good sir
Sir loop checking kaise hota hai aur kya kya instrument use hota hai. ...please
V nice
Sir mujhe hydrolic pressor transmitter ki calibration Karni h working pressure 19mpa please sir tel me
Sir,
First, Thanks for this nice and helpful tutorial, also, kindly use English if it is possible, to
make your videos useful for every one around the globe.
Second, even if you don't use any 1-5 VDC measuring device here, you need to have a
loop resistor in order to maintain the communication between the transmitter
and the HART communicator. If you bypass the resistor, the power supply filter capacitors
will eat your HART signals (AC)..!
[By an instrumentation infant and an electronics teen :) ]
Thank You Prajeesh Dev Poonkulanji Sir,
For Your Valuable Suggestion and Support Thank You.
absolutely correct sir..........
You tell that to use English language but some peoples are not comfortable in English
visit ua-cam.com/channels/d6LP1BncR9AZbPf9-dLNCA.html
Yyyyyyyyyyyyyyyyyyyyyyyyyyyyp
Sir normal range 0 to 10 psi rehta he na to. Apne kyo 2 To 10 psi bataya
Perfect
Sir pls upload pt ft calibration procedure and loop checking off these instruments
Sir. ..please explain why we used 0-24 v dc for this calibration
Thanks you for the nice vidoe
Please let me know Equaliser lime to br open or close
Your Welcome, Mohammad A ji
Mr. mohammad equaluser vavle will be close here if u will not close equaliser then ur given pressure through hp line is also will also go lp line through equliser nd u find zero pressure in output
Sir ICLX valves k bare me btae plz
Sir, where we will connect multi meter probs to read the 4 to 20 milli amps current value.
plz reply me
I am waiting for your reply
You can connect Multi meter to Transmitter and power supply
visit ua-cam.com/channels/d6LP1BncR9AZbPf9-dLNCA.html
10 minutes ka video banaya karo sab dekhe ga
cabel Meghering kya hota hai or kise ...!
Hello sir aapne .analog card bataviya ..digital card ne 1to 5 v dc opert
Sir pls discrib what is mean by maximum working pressure and callibration rang in transmitter which is written on transmitter description plat
Maximum pressure is supposed 10 bar.....and calibration range is 0 to 6 bar.....it means your tx is calibrated 0 to 6 bar....if you want calibrate this tx from 0 to 10 bar you can bcoz maximum pressure is 10 bar it's diaphragm capacity
very good thank you
Sir instrument technician course online kara sakte hai
Sir multimeter ko kaise series me lagaye ge
Nice video
Thanks Nadeem Khan Ji
Sir Endress hauser 50 flowmeter calibration process explain plz in hindi main
Ashok hart communicator series m lagta h
Hart se MA K SETTING KASE KARTAY CAN YOU SHOW ME VIDEO
Please show us prectically 20 ma and 4 ma testing
And how to connect 250 ohm resistance in the wire
We r waiting for ur kind response
Connect 250 ohm resistance between common & analog input signal. you can connect either your controller side or DPT side.
Sir plz make a video about what is mean by maximum working pressure of transmitter
Maximum working pressure is suppose 10 bar .....10 bar means it is a full capacity of diaphragm
Sir TT, FT And LT ka bhi Calibration ka vedio banayye....
OK, Shubham Ji Hum bahut Jalt TT, FT and LT par bhi video banayenge Thank You
visit ua-cam.com/channels/d6LP1BncR9AZbPf9-dLNCA.html
PT ka calibration batao Bhai..kaise karege
Good
Thanks, Pradeep Singh Ji
sir DP transmitter se level measurement kaise krte hai iske bare btaiye
How to calibrate a dpt -200 to 200 mmwc
Sir. Dp type level transmitter calibration ka video banaye
Sir pls give some basic knowledge about somatic manager programing
Dp gauge cailbretion prefect
Sir plz discrib about PID controller and how we do PID tuning. sir plz video make in hidi.
visit ua-cam.com/channels/d6LP1BncR9AZbPf9-dLNCA.html
Very very thanks
Love u sir
Nice
Thanks, Manish Parab Ji
Sir please make a video on control valve covering all info about thay
OK, Vishal Yadav Ji
Vaise aapki qualification kya h
iam from bangladesh
these procedure applicable for Foundation Fildbus DPT
visit ua-cam.com/channels/d6LP1BncR9AZbPf9-dLNCA.html
sir , plz plz plz dp switch & dp guage Ka calibration video banaiye
OK, Jitendra ji hum iske liye bahut jald video banayenge, Thanks for your Valuable suggestion.
thanks sir g
Sir
LT k callibration ka video banaiye
Bena pressure deye hum chak karle ge😂😂
Sir ,please make video on HART function
visit ua-cam.com/channels/d6LP1BncR9AZbPf9-dLNCA.html
Sir long video banaye
His me ewulizevalve kyabkarega
Sir evarina TELUGU lo chyppandi Industrial Instrumentation Instruments Gurenchi....