  • @ebayscopeman
    @ebayscopeman 9 років тому +1

    To increase the accuracy of the measurement when doing the calibration the calibration line should be snapped to the same edge of the calibration scale. That is pick either the left edge or the right edge but make sure you pick the same side both times. You could of course pick the center but this is usually harder to define.

  • @rogerlaurel8202
    @rogerlaurel8202 9 років тому +6

    Is all of this available for the Apple software? I don't see these features!

  • @JET8400
    @JET8400 7 місяців тому +2

    You should calibrate to the longest distance you can see for better accuracy and use the cert results and not nominal to input as the actual. If you cal to .05 you have used your scale six times at .3mm. If you cal to 1, you’re at .3 times. Then you would measure .30 at .3 and not .28. Thanks for the info. Cheers!

  • @jeff35372
    @jeff35372 9 років тому +1

    Great and Thanks for putting this on Y/T.
    I have another set of Objectives and i would like to calibrate the as well, How can I add a 60x or even another list for a different set of Objectives without losing my current settings?
    MU500 version x86, 3.7.4460

  • @GrayPlayer
    @GrayPlayer 9 років тому +1

    Have calibrated objectives. Have question explaining at what point the camera actually clicks the "shutter" compiling images for stacking.

  • @unnitamahato1579
    @unnitamahato1579 6 місяців тому

    Thank you so very much... What I was looking for , this id thf video..❤

  • @johnmarshall124
    @johnmarshall124 8 років тому

    Hi,At 16:18min. You have calibrated at 0.05mm, you then measure to the 0.30 mm line but the software states it is 0.28mm. Sorry but I do not understand. The long lines are 0.1 mm apart, your line is three lines long so it should measure as 0.30mm not 0.28mm.RegardsJohn

  • @MarcusCorpa
    @MarcusCorpa 6 місяців тому

    I am using a mac MU1003 and amlite software. I can get to the calibration box but the buttons are inactive.

  • @kamelpalumbo
    @kamelpalumbo 7 років тому

    Can i calibrate the microscope using a neubauer chamber????. Thanks in advance!

  • @kaveh341
    @kaveh341 9 років тому +1

    I'm having trouble with mine. Every time I hit calibrate, whether I measure before or after closing and reopening the camera, it measures wrong. I followed all the steps here and I'm using windows 7 if that's a factor

  • @eismaildds1
    @eismaildds1 6 років тому

    I can't connect my AmScope to my mac! What soft ware and where to get it from to make it work? The cam scope doesn't appear to be connecting to my mac, please help! I installed the program from the disc, then downloaded it online, but the scope never appeared on the app!!

  • @kinito1
    @kinito1 7 років тому

    I have a MU1803, my gemology microscope is in focus, but the display in the software just won't focus no matter what I do.

  • @maycol8950
    @maycol8950 9 років тому

    How to i calibrate in lower objective on a stereoscope there is only 4, 10, 20, 40, 100 objectives

  • @chiliorozco
    @chiliorozco 8 років тому

    hi. I have a MU300 but when I put it on the 3rd lence all becomes black. I already checked and I put out all the covers from the camera lens, and still like this cannot see anything to do the calibration. Any help please?

    • @AmScopeVideos
      @AmScopeVideos  8 років тому

      Are you using one of our SM microscopes? If so check to see if there is a silver pin on the left hand side and pull it to the outer position to open up the trinocular port. If not, then please contact our customer service line or shoot us an email through our main website so that we may assist. Thank you!

    • @chiliorozco
      @chiliorozco 8 років тому

      Yes!!! It was that :) thanks

    • @AmScopeVideos
      @AmScopeVideos  8 років тому

      You're welcome, enjoy your weekend!

  • @nunzioguaragno9384
    @nunzioguaragno9384 6 місяців тому

    HI. I can't do the calibration. The ruler doesn't appear to me, only the image that the telescope sees. how to solve it, thanks

    • @LarsLarsen77
      @LarsLarsen77 15 днів тому

      You need a calibration slide.

  • @Zstray
    @Zstray 9 років тому

    Thanks!

  • @FernandaRocha-vv2ps
    @FernandaRocha-vv2ps 6 років тому

    Alguém no Brasil sabe utilizar este programa?

  • @evazupan4214
    @evazupan4214 7 років тому

    How can I calibrate md35 camera?

  • @chiliorozco
    @chiliorozco 8 років тому

    and the dropbox link doesnt work

    • @AmScopeVideos
      @AmScopeVideos  8 років тому +1

      Software link has been updated - it now leads directly to our website which has the most up to date version available.

  • @saba500
    @saba500 7 років тому +1

    I Need that calibration scale too.

    • @AmScopeVideos
      @AmScopeVideos  7 років тому

      The calibration slide can be purchased here: www.amscope.com/microscope-stage-calibration-slide-for-usb-camera-0-01mm-stage-micrometer.html

  • @pfrpadilla
    @pfrpadilla 6 років тому

    too long video for such a simple thing ...