    How To Develop Terminal Sterilization Cycle?
    Why Air Removal not required in Terminal Sterilization?
    Why Vacuum not required in Terminal Sterilization?
    What is D Value?
    What is F0 Value based Sterilization Cycle?

  • @kamalkishorpant9720
    @kamalkishorpant9720 Рік тому +1

    Many many thanks Sir for teaching , gratitude 🙏

  • @nitinkmahajanmahajan223
    @nitinkmahajanmahajan223 7 місяців тому +1

    Nice video

  • @tarunthakur2009
    @tarunthakur2009 Рік тому +1

    Many thanks sir

    • @PHARMAVEN
      @PHARMAVEN  Рік тому

      Thank You For Your Appreciation, Kindly also share to your friends and groups and juniors 🎉🙏👍

  • @bhupendrashukla1972
    @bhupendrashukla1972 Рік тому +1

    Useful

  • @nirmalsartana4061
    @nirmalsartana4061 Рік тому +1

    Very nice sir

  • @Helmet_red
    @Helmet_red Рік тому +1

    Sir could you please explain about the ventilator cycle in brief please

  • @pukhrajmaske7047
    @pukhrajmaske7047 Рік тому +1

    Sir hindi me sabhi video bano

  • @BrandBabuSoni
    @BrandBabuSoni Рік тому +1

    Hindi me bhi bana do ek video

  • @ManishKumar-fp3wn
    @ManishKumar-fp3wn 3 місяці тому

    Which guidelines refer terminal sterilization.???

  • @Animal-m5y
    @Animal-m5y 9 місяців тому

    How to detrmine fo value using bacilus subtilus at 108 degree for 60 min

  • @md.shahidul9013
    @md.shahidul9013 Рік тому +1

    Nice,My question,what’s area is required for terminal sterilizer,is non class area okay? Please give answers with reference

    • @PHARMAVEN
      @PHARMAVEN  Рік тому

      After filling and sealing, the product is basically finished product and doesn't need to store in classified area. For terminal stabilizer, CNC (Controlled Non Classified) area is acceptable

  • @kalyanvenkatesh9523
    @kalyanvenkatesh9523 Рік тому

    F0 value development?

  • @AnilSingh-hg5pz
    @AnilSingh-hg5pz Рік тому +1

    How vials are kept inverted or upright during ts

    • @PHARMAVEN
      @PHARMAVEN  Рік тому

      Upright only during terminal Sterilization

  • @bholuyadav9276
    @bholuyadav9276 Рік тому +1

    Which guidelines is for terminal sterilization please reply

    • @PHARMAVEN
      @PHARMAVEN  Рік тому

      European guide: Guideline on the Sterilisation of the medicinal product, active substance, excipient and primary container
      USFDA: Guidance for Industry Submission of Documentation in Applications for Parametric Release of Human and Veterinary Drug Products Terminally Sterilized By Moist Heat Process

  • @niravbhatt1473
    @niravbhatt1473 Рік тому

    Thanks sir
    Temperature mapping required in what size containers?

  • @marwanfahmy8608
    @marwanfahmy8608 Рік тому +1

    If i use over kill approach of 8log reduction why i consider 121/10min cycle not a Sterilization cycle.and what is the minimum time parameters if I need to change for higher temperature as 134.knowing that f0 is12.... thanks.

    • @PHARMAVEN
      @PHARMAVEN  Рік тому

      Overkill depends upon D value of product, for example if D value is 1, then 121 degree C at 10 minutes will give 10 log reduction, Overkill means 12 log reduction.

  • @kalyanvenkatesh9523
    @kalyanvenkatesh9523 Рік тому

    Who will fix F0 value sir.