  • @TestTest-vv6vh
    @TestTest-vv6vh 21 день тому

    Is getting the background conductivity just zeroing the device or is that done differently

  • @soorishadab
    @soorishadab 6 місяців тому

    Acceptance criteria of solt in industry ?

  • @johnbasford2529
    @johnbasford2529 6 років тому +1

    I am curious to know how its possible to flush clean the resuable possipatch between uses - don't you get the risk of soluble salt hang up in the patch?

    • @DeFelsko
      @DeFelsko  6 років тому +6

      Hello John,
      Thanks for your interest in our PosiTest SST instrument and reusable PosiPatch.
      Cleaning the PosiPatch between tests is as easy as rinsing it thoroughly with DI water. Extensive testing has shown that this simple rinse is sufficient to remove soluble salt residue from the PosiPatch. If higher than expected results are obtained, use a new PosiPatch to verify results.
      Please feel free to contact us to discuss this further +1-315-393-4450. It is our pleasure to assist you.
      -Terry

  • @instinct0199
    @instinct0199 3 роки тому

    Hey Guyz is distilled water also possible to use?

    • @DeFelsko
      @DeFelsko  3 роки тому +1

      Like all soluble salt instruments that test in accordance with ISO 8502-6 (the Bresle Method) and ISO 8502-9, the PosiTector SST requires the use of DI (deionized) water to obtain a valid reading. Please contact us at techsale@defelsko.com if you require additional assistance.