  • @TheJoshtheboss
    @TheJoshtheboss Рік тому

    Just wondering why does the cell need to be removed for the intensity test? I don't understand what is the reasoning behind the fact that it passes only when the cell out?
    Since when analysing the cell will be in obstructing some of that light?

    • @Spacefish007
      @Spacefish007 Рік тому +1

      The Test is intended to check the detector/lamp itself and not the cell, there is a separate "Cell test" for that, which meassures the ratio of light the cell itself is absorbing.
      Otherwise the cell would just introduce unknowns, like solvent / sample or contaminents left in the cell or cell windows.