HPLC Maintenance - Replacing the Pump Seals on an Agilent 1100/1200/1260 HPLC Pump

Поділитися
Вставка
  • Опубліковано 22 сер 2024
  • HPLC maintenance tutorial on how to replace the seals on the pump head on an 1100/1200/1260 Agilent HPLC.
    For more information, please visit www.Agilent.com.

КОМЕНТАРІ • 9

  • @spencerscott7826
    @spencerscott7826 2 роки тому +1

    You don't mention replacement of the film washers. Are these no longer being used? Recommended? Thank you

  • @djcfrompt
    @djcfrompt 3 роки тому +2

    Do you have a similar video for a pump with active seal wash?

  • @ka-zx6bb
    @ka-zx6bb 5 років тому +5

    What Agilent part no. are these seals? Thank you for the info.

  • @pdjobanputtra
    @pdjobanputtra 4 роки тому

    Thank you

  • @dzfz2100
    @dzfz2100 7 років тому +2

    How often does this need to be done?

    • @agilent
      @agilent  7 років тому +4

      Replacing the pump seals is a part of routine maintenance or a Preventative Maintenance(PM). Pump seal wear depends on usage, how often the pump is used and what solvents are being used. Agilent recommends replacing the pump seals once a year as a general rule.

    • @dzfz2100
      @dzfz2100 7 років тому +6

      Brilliant - thanks so much! I've taken over care of our Agilent GPC system from a former postdoc, and his answer to my question "when did you replace the seals in the last 2 years" was "never". So I guess I'll be buying some new seals!

  • @Technicalsolutions3536
    @Technicalsolutions3536 5 місяців тому

    Please share in india supplier