Neptune Trident Setup and the "No water sample" issue 4K

Поділитися
Вставка
  • Опубліковано 2 лют 2025

КОМЕНТАРІ • 31

  • @NerdistAquarist
    @NerdistAquarist 2 роки тому +3

    Informative video! I've had the no water sample for two other reasons as well.
    1 - Dirty mixing chamber. Water was filling, but wasn't being read by the optical sensor. I cleaned the chamber with a Q-tip.
    2 - Detritus buildup between solenoid 3 and the manifold, causing clog.

  • @gibbnal
    @gibbnal 3 роки тому +4

    Thanks for posting this. One additional step if both waste line and sample line are clear and you still have problems. The black manifold can potentially clog. If you remove all of the tubes, you can flush it by carefully syringing warm water through each port. You can selectively close them with your fingers to increase the volume that flushes through individual ports. It's fiddly. Note the orientation of the ports before disassembly. The 3 reagent ports are larger diameter than sample and waste. The small diameters can certainly clog (mine took 18 months) but I believe they are necessary for the precise droplet size required in such a small test volume. Some sort of prefilter would help, but you will still have biofilm to deal with eventually.

    • @rahimsach
      @rahimsach 4 місяці тому

      That's a super helpful comment. I had cleaned the manifold but re-connected it wrong, I thought the thicker ports connected to the sample and drain., no wonder it didn't work.

  • @jessetilden4334
    @jessetilden4334 2 роки тому +1

    Thanks for this, man! Worked like a charm.

  • @muddlovin
    @muddlovin Рік тому

    just had the same issue come up on mine. Thanks for the information!!

  • @steveroberts9331
    @steveroberts9331 2 роки тому

    Thanks dude - helped me fix my no water sample error.

  • @nautiswerk
    @nautiswerk 3 роки тому +1

    Thank you for posting this! I had this right out of the box, too. The rubber lines obviously need some rethinking.

    • @PurelyAquatic
      @PurelyAquatic  3 роки тому +1

      Agreed, those rubber lines are way too small.

  • @geraudk
    @geraudk 3 роки тому +1

    Great video, thanks for posting. This helped me unclog my water intake, as well as the manifold itself (using a tiny sewing needle)

  • @TheCluelessReefer
    @TheCluelessReefer 2 роки тому

    Thanks for posting, just installed mine only to have the Sample line pinched/kinked. Used the needle syringe from a Red Sea test kit and couldn't get any water to push through, and I visibly could see a pinch... I tried to massage it out but didn't work.

    • @PurelyAquatic
      @PurelyAquatic  2 роки тому

      If its still not working for you, give the tech support guys a call. Really helpful team!

  • @clovercompetitions4714
    @clovercompetitions4714 2 роки тому +1

    Great video exactly what was wrong with mine 🙏🏻

  • @shoxite
    @shoxite 3 роки тому

    Damn dude I just got my Trident today. Having the same issue 😂. What can we do besides laugh? Lol gonna go through this video again and follow your steps. Thank you and I appreciate the video! Seems like many people are having the same issue.

  • @leongorani5121
    @leongorani5121 2 роки тому

    Would this fix also work for the “Test A Fail” message, it just came up in my trident. Thanks

  • @boblivingston4841
    @boblivingston4841 Рік тому

    I had the same thing happen to one that I had for 1 month.

  • @phillandau350
    @phillandau350 3 роки тому +1

    Did you flush only the tube connected to #3 or did you use the syringe to flush water through #4 also?

    • @PurelyAquatic
      @PurelyAquatic  3 роки тому +1

      I used the syringe to flush water on both lines 3 and 4.

  • @vincenthong9536
    @vincenthong9536 2 роки тому

    why my trident still show up is O?

  • @abitsalty7603
    @abitsalty7603 3 роки тому +1

    You saved me the aggravation of calling the creators of cheaply built contraption. Thanks

  • @ThePthinker
    @ThePthinker 3 роки тому +2

    guess im in the same boat...Not good for such an expensive piece of equipment

  • @ramezriad4047
    @ramezriad4047 2 роки тому

    There’s only one pump in the Trident; what you call pumps are solenoid valves that are all connected to a manifold in the center

  • @juanmorales3577
    @juanmorales3577 3 роки тому

    Yeah I have that issue today how do I do to fix it?

  • @MW-ut3dk
    @MW-ut3dk 3 роки тому +6

    For how much you pay for this equipment it shouldn’t be clogged by any particles considering it’s brand new. Are these things qc’d before leaving the factory? They should be tested before shipping out. Consumers shouldn’t need to take apart new equipment and repair them… wondering if it’s this easy to clog these feed lines then why didn’t they use larger diameters? I know, less reagent used per test, but having to disassemble this frequently to maintain it doesn’t sound like fun. Alternatively they could’ve increased concentration of reagents to deal with larger sample size, but it’s just me going down the rabbit hole🤔. Just seems too finicky to be reliable, but let us know about the real world owner’s experience.

    • @PurelyAquatic
      @PurelyAquatic  3 роки тому

      I agree the lines should be larger in diameter. Probably would mean they would have to get a bigger pump/solenoid, which would mark up the price even more.

    • @MW-ut3dk
      @MW-ut3dk 3 роки тому

      @@PurelyAquatic If jebao can give you four heads for less than $100 usd retail then it wouldn’t cost much more😜

    • @NAZ576
      @NAZ576 3 роки тому +1

      I have the same problem I will try to do same you, hope it works with me 🙏

    • @PurelyAquatic
      @PurelyAquatic  3 роки тому +1

      @@NAZ576 If it still doesn't work you can call their tech support

    • @NAZ576
      @NAZ576 3 роки тому +1

      @@PurelyAquatic it’s working 🤩🤩 thanks