Question on ps-cal. When checking low power (on a E9304A sensor). PS-Cal asks is 20 dB Attenuator factors are included in the mount factor, I reply with YES. But then PS-Cal asks for the 20 dB value. What is this for, what do I enter? I tried enter the stamped value on the Attenuator (21.5) and 20 dB. I current issue is, the senor -20 readings are reading low -20.1 dBm (when 20 was entered). And -23 dBm when 21.5 was used.
Question on ps-cal.
When checking low power (on a E9304A sensor). PS-Cal asks is 20 dB Attenuator factors are included in the mount factor, I reply with YES. But then PS-Cal asks for the 20 dB value. What is this for, what do I enter?
I tried enter the stamped value on the Attenuator (21.5) and 20 dB.
I current issue is, the senor -20 readings are reading low -20.1 dBm (when 20 was entered). And -23 dBm when 21.5 was used.