Cobas 6000- Changing the Photometer Lamp

Поділитися
Вставка
  • Опубліковано 3 гру 2024

КОМЕНТАРІ • 12

  • @Isra_Saberhagen
    @Isra_Saberhagen Рік тому

    Es justo lo que hice hoy! Lástima que encontré el video muy tardé, sin embargo todo está en el manual 🎉

  • @alvisgineika9306
    @alvisgineika9306 5 років тому

    Very Interesting and seams a lot better than many other devices

  • @calamariaxo
    @calamariaxo 9 років тому +2

    Even though Roche recommends changing it every month we've successfully run lamps for almost three months without errors. Just remember to monitor the photometer check each day so it doesn't suddenly blow out on you. There's money to be saved in at least delaying it a bit :)

    • @robertbounds8112
      @robertbounds8112  9 років тому

      +Peter Cruickshanks Yes, it is true. Roche recommends the lamp be changed monthly but they typically last longer than that (in my experience). I've never been bold enough to wait three months though.

    • @calamariaxo
      @calamariaxo 8 років тому +1

      +Robert Bounds Very understandable, lets just say our longest run on one lamp was not quite intended. Props for the videos/channel btw, too few labtech things out there :)

    • @robertbounds8112
      @robertbounds8112  8 років тому +4

      +Peter Cruickshanks Thank you. I was thinking about making an overview video of the profession for people who don't know what it is we do. Nobody knows we exist, lol.

  • @efsunderler8386
    @efsunderler8386 4 місяці тому

    After this process, the device gives an "abnormal cell blank difference" warning. What's the solution?

  • @dinosaur14jb
    @dinosaur14jb 2 роки тому

    I was trained to let the lamp sit for 3 hours before turning power back on, is that not necessary?

    • @robertbounds8112
      @robertbounds8112  2 роки тому

      I’m pretty sure the recommended procedure is 30 minutes before using it. They also suggest 30 minutes before removing it so it has time to cool.

  • @walidtawfik1289
    @walidtawfik1289 2 роки тому

    Pls.. Machine give message abnormal temperature and i hear sound behind machine