Calibration adjustment of a Transcell TI500E

Поділитися
Вставка
  • Опубліковано 15 гру 2024

КОМЕНТАРІ • 7

  • @PhilSet-10
    @PhilSet-10 4 дні тому

    I’m a little confused about your “400” entry in F17. I thought that once you zero the scale in F16 that you do it with an empty scale at which point you go to F17 and add a test weight onto the scale which should then appear on your scale. You went straight from clearing F16 to going into F17 without adding a test weight. What exactly was the “400” for that you entered in F17?

    • @Raddamez1215
      @Raddamez1215  4 дні тому

      F16 is a zero adjustment F17 is a span adjustment when you enter F16 you zero out the raw count press enter and then at F17 you punch in the weight apply your weight press enter then power down the indicator and switch the switch back to its original position

  • @Scaleman
    @Scaleman Рік тому

    These are helpful thanks!

  • @kuldeepsinghyadav8783
    @kuldeepsinghyadav8783 Місяць тому

    The scale display not coming to main display 0000, it's directly show CH 0, how we can go to main display ?? Transcell TI-1500B

  • @jaimeevillalobos6465
    @jaimeevillalobos6465 Рік тому

    Error 1 /2 /3
    how can i fix these errors??