    Do you know anything about the error messages ie. Err03. I have been getting them lately and cant figure out what they mean. Thanks

    • @MaxtecO2
      @MaxtecO2  2 роки тому +2

      Hi M, thanks for reaching outl. The E03 error code means "there is no valid calibration data available". Make sure unit has reached thermal equilibrium & perform calibration as described in the user manual. You can find a digital version of our IFU here: www.maxtec.com/product/delivery/mixers-blenders/maxventuri/. If the problem still persists, submit a claim here: www.maxtec.com/technical-support/. For any other questions or concerns, please message customerservice@maxtec.com. Thank you!