Medtronic Autolog Cell Saver Tutorial

Поділитися
Вставка
  • Опубліковано 1 жов 2024
  • 00:06 Introduction
    00:21 Gathering the required disposables
    00:37 Properly documenting pertinent information
    01:22 Opening and installing the reservoir
    02:07 Opening the wash kit and installing the waste bag
    02:47 Installing the wash kit
    06:08 Installing the blood filter and checking connections
    07:08 Installing the vacuum line
    07:31 Final setup and covering of the cell saver for later use
    08:12 Spiking the wash solution
    08:34 Turning on the machine and vacuum
    08:48 Priming the cell saver
    08:57 Hooking up the table pack lines to the cell saver
    09:17 Coating the reservoir filter
    09:30 Adjusting the heparin drip
    09:43 Processing reservoir volume
    10:33 Running the final cycle

КОМЕНТАРІ • 5

  • @meshellwhite8994
    @meshellwhite8994 Рік тому +1

    Where can I get the blue tool to push the tubing in

  • @HeartPumper
    @HeartPumper 8 років тому +1

    Handy tip, in case of blood getting *somehow* out of a washing bowl. Press On together with green button, hold it to the moment of screen shows software version and blinking led by green bttn. Pour lukewarm water with soap to the compartment holding washing bowl. But before it, lay some 'hydrophilic' mats or other material, under the device. Repeat and flush with clean water.

  • @TheMrriyadh07
    @TheMrriyadh07 7 років тому +1

    How much heparin exactly?

    • @robertlawson1184
      @robertlawson1184 3 роки тому +3

      30ml per 1,000 saline bag.
      15ml (15,000) per 500 saline. Happy 4 years answer.