STATIM 2000 Fault 16 Water Pump Replacement

Поділитися
Вставка
  • Опубліковано 5 жов 2024
  • Fault 16 could be many things, but I played the odds. Could have just taken the pump out and cleaned the screens, both inlet and outlet but that may not have worked. Bought a pump from a reputable seller on Ebay that came out of a machine with a different issue. Easy to do.

КОМЕНТАРІ • 3

  • @Samlol23_drrich
    @Samlol23_drrich  3 місяці тому

    Not too many videos of “real people” (me) trying to fix something they haven’t before,so I figured I’d make this one. Are most dentists just gonna call someone and spend $1000 or more when they get an error code? Probably, but that’s not me. I enjoy the challenge of repairing my own things, if I can. In reality, this was no more than a few screws to swap out the pump. I suggest you download the service manual first, as there are 3 types of water pumps, and the tubing can be of a different diameter that affects how much water the pump is transferring to the boiler.

  • @alexthehandyman3494
    @alexthehandyman3494 2 місяці тому

    Hey Rich, If that didn't work you could have taken the pump from your wet saw. Pretty much the same. JK

    • @Samlol23_drrich
      @Samlol23_drrich  2 місяці тому

      @@alexthehandyman3494 thanks. There are a lot of similar ones. I changed a pump in my other sterilizer made by a company that supplies pumps for espresso machines. It cost $40 directly but it would take 2 weeks to arrive from Canada so I had to spend $130 and buy it from a dental supplier. Crazy.