Fluke Multimeter calibration

Поділитися
Вставка
  • Опубліковано 22 лис 2018
  • a simple demonstration on how a multimeter is calibrated using a multiproduct calibrator (Fluke 5522A)
    Click below link to see the full procedure..
    bit.ly/multimetercalibration

КОМЕНТАРІ • 38

  • @Jan-cm2ye
    @Jan-cm2ye 9 днів тому

    Today I learned what calibration actually means. Guess after calibration I'm making adjustments from now on.

  • @marcjj4438
    @marcjj4438 3 роки тому +9

    This is not calibration, this is only verifiaction of accuracy, if the accuracy is not in range and need to be adjusted then you do the calibration, when you correct the measure of the DMM to have a better accuracy then that is the calibration. You can find calibration manual for some Fluke DMM online, you could do it yourself but you need a proper equipment and sometimes for work you will need a certificate of calibration, so you have to paid the calibration to authorized labs.

    • @calibrationawareness7956
      @calibrationawareness7956  3 роки тому +3

      Calibration and verification are the 2 confusing terms when implementing the calibration process. During calibration, our main purpose is to determine the accuracy of our instruments, this is done through comparison of readings to obtain the error (or to be complete, bias and measurement uncertainty).
      During verification, our purpose is to determine if the result of calibration has ‘passed’ or ‘failed’ based on a given tolerance limit. From these definitions, you can now determine your purpose if to ‘calibrate’ or to ‘verify’. Read more in this link>> calibrationawareness.com/how-to-differentiate-calibration-verification-and-validation
      Thanks for the comments.

    • @joshuadavid5147
      @joshuadavid5147 2 роки тому +2

      Calibration is just comparing, not adjustment.

    • @SteverRob
      @SteverRob 2 роки тому

      Nope

    • @ariadarabi
      @ariadarabi 2 роки тому +1

      @@calibrationawareness7956 Interesting. It looks like I have learned something today. Thank you!

  • @itsonlyme9938
    @itsonlyme9938 2 роки тому

    I have a 177 fluke on ma is reading about 3ma low compared to the amps range not if this is correct ?

  • @gere364
    @gere364 9 місяців тому

    It is possible to do correction error adjustment in the Fluke 115 multimeter?

  • @JOHNY20103
    @JOHNY20103 5 років тому

    I Have a problem in my 5520 .it shows error 1503(current lim exceeded )fter 3.3v(DC).CAN ANYONE HELP ME

  • @hermanvisser4034
    @hermanvisser4034 3 роки тому +1

    Can I ask if the Fluke multimeter can be adjusted if it is out of calibration?

    • @calibrationawareness7956
      @calibrationawareness7956  2 роки тому

      Yes, it can be, some models have service instructions for adjustment provided by the manufacturer that we can follow. Just look for the service manual.

  • @user-uv9vx4fn3r
    @user-uv9vx4fn3r 7 місяців тому

    Hello! Можешь показать как получить гармоники по напряжению и по току?

  • @Rajeev1200
    @Rajeev1200 2 роки тому

    Dear Edwin sir, I have a question,
    How will we use our std. error (which we got in std. calibration certificate) during this calibration ?

    • @calibrationawareness7956
      @calibrationawareness7956  2 роки тому

      Hi Rajeev,
      Std error is used as a correction to compensate for the error encountered. For every error that you have, a corresponding correction factor is calculated (which is the opposite of the error).
      Check out my article in this post where I presented how to use the correction factor in calibration results. >> calibrationawareness.com/simple-ways-to-apply-correction-factors-in-a-calibration-certificate-if-the-exact-value-you-need-is-not-given

  • @daixtr
    @daixtr 3 роки тому

    Do we need special cable in order to calibrate??

    • @calibrationawareness7956
      @calibrationawareness7956  2 роки тому

      Yes, there is a special cable designed for calibration with good insulation, low resistance value, and can withstand high current range.

  • @justineodal754
    @justineodal754 2 роки тому +2

    sir what if off yung reading ng meter compared sa calibrator, may paraan ba na ma adjust ang calibration ng fluke dmm? meron ako fluke 28ii na medyo below (200-300mv in DC, less 1v or more in AC) ang reading nya compared sa aking fluke 287. i wonder kung pwede ba ma adjust yung reading ng unit at least to its specification. Im aware na mas accurate ang 287 as per specification pero dapat sana halos magkapareho lang readout nila at least in first decimal. tia

    • @calibrationawareness7956
      @calibrationawareness7956  2 роки тому

      hindi ako sure kung pwede sa model na yan pero ang pagkakaalam ko basta Fluke pwede sya. karaniwan nasa user's manual kung paano iadjust. meron ka lang pipihitin na trimpot para maadjust sya. pwede mo gamitin na reference yung mas accurate na dmm mo. Salamat sa pag view ng video.

  • @PAR1171
    @PAR1171 3 роки тому +1

    How calibrate digital multimeter in measure mode .

    • @calibrationawareness7956
      @calibrationawareness7956  3 роки тому +3

      A multimeter is always in a measure mode, it is receiving a signal from the calibrator.

    • @leonordeborbon2121
      @leonordeborbon2121 Рік тому

      internal board have some screws there thar you turn +/- to center bias of several parameters

  • @wahidhamraoui130
    @wahidhamraoui130 4 роки тому +2

    Calibration. Where is it !!!!????

  • @tumaliuanjames3788
    @tumaliuanjames3788 3 роки тому

    Sa pilipinas ba to boss magkano magpa calibrate fluke 177

  • @edergomez4396
    @edergomez4396 3 роки тому +5

    this is just a verification, not calibration.

    • @calibrationawareness7956
      @calibrationawareness7956  3 роки тому +2

      Calibration and verification are the 2 confusing terms when implementing the calibration process. During calibration, our main purpose is to determine the accuracy of our instruments, this is done through comparison of readings to obtain the error (or to be complete, bias and measurement uncertainty).
      During verification, our purpose is to determine if the result of calibration has ‘passed’ or ‘failed’ based on a given tolerance limit. From these definitions, you can now determine your purpose if to ‘calibrate’ or to ‘verify’. Read more in this link>> calibrationawareness.com/how-to-differentiate-calibration-verification-and-validation

    • @SteverRob
      @SteverRob 2 роки тому +1

      Adjustment to nominal can be a part of a calibration, but not always. If the UUT is within manufacturer's specs when compared to a lab standard (usually a 4:1 accuracy ratio) it is calibrated.

  • @tyketajonera6336
    @tyketajonera6336 5 місяців тому

    You need to explain it more.

  • @filipinopatriot9202
    @filipinopatriot9202 2 роки тому +4

    Calibration means you need to adjust the equipment. Verification is definitely not a calibration.

    • @user-bg9rc1sx1s
      @user-bg9rc1sx1s 2 роки тому +1

      Calibration is a testing against a good standard per calibration manual. Adjustment is not a calibration, but can be called a cal/adjust in some cases

    • @SteverRob
      @SteverRob 2 роки тому

      incorrect.

  • @kfcmcdonalds
    @kfcmcdonalds Рік тому

    Question is : how did you CALIBRATE the meter and where's the calibration control on the meter ?
    You've only showed various readings on an already calibrated meter that's all.

    • @leonordeborbon2121
      @leonordeborbon2121 Рік тому

      usually you have to expose the board and find some screws there that you can turn +/- the central bias of several parameters

  • @AlainHubert
    @AlainHubert 2 роки тому +1

    This multimeter is definitely NOT calibrated properly. It should read 250VDC and 250 VAC, not 249.8VDC and 249.8VAC. This is only a verification, not a calibration.

    • @calibrationawareness7956
      @calibrationawareness7956  2 роки тому +2

      First of all, this is just a simple demonstration. Calibration may or may not include adjustment. The main point in calibration is the comparison of the std reading with the DUT reading to determine the error. Read more in this guide. Thanks... >> calibrationawareness.com/how-to-differentiate-calibration-verification-and-validation

    • @user-bg9rc1sx1s
      @user-bg9rc1sx1s 2 роки тому +1

      You definitely don't understand what the calibration is if you never heard about uncertainty 🤦‍♂️