We have changed several Mindray lamps but the reaction temperature keeps fluctuating (reaction temperature out of range) and light intensity too weak.... Please what should I do to resolve this challenge, because the analyzer is not working right now... Please Sir,I awaits your quick responses . THANKS sir.
Both things r different. 1. Reaction temperature out of range depends on its heater and sensor). That u have to verify. 2. Light intensity too low. U can adjust filters by using debug software.
Sir when we are doing in qc calibration.first we have to do a blank test(water+reagent).then standard (standard +reagent) . Is it sir? Because I have a doubt.
Together u will do. Keep your calibration rule as 2 point. Then run calibration. It will take blank as well as calibrator. Then u can run QC. Although u can run qc same time but it would be better to do it later.
Actually Mindray or any brand Semi Automated Chemistry Analyzer is not allowed in GCC countries. That’s why I can’t but still i will try if i will get a chance.
Please Sir,how can we resolve the problem alarm "light intensity too weak and the reaction temperature is always out of range"??
Hi,thanks for the video but how do i replace a lamp
Can you do video about calibration
My device shows low light intensity error. What should I do?
Change lamp.
Difference with Bs 230?
We have changed several Mindray lamps but the reaction temperature keeps fluctuating (reaction temperature out of range) and light intensity too weak....
Please what should I do to resolve this challenge, because the analyzer is not working right now...
Please Sir,I awaits your quick responses .
THANKS sir.
Both things r different.
1. Reaction temperature out of range depends on its heater and sensor). That u have to verify.
2. Light intensity too low. U can adjust filters by using debug software.
Sir how to do it please make a video. This error happens here also.@@biomedicalengineeringbysha5215
How to change the lamp
How can I insert the control values?
Sir when we are doing in qc calibration.first we have to do a blank test(water+reagent).then standard (standard +reagent) . Is it sir? Because I have a doubt.
Together u will do. Keep your calibration rule as 2 point. Then run calibration. It will take blank as well as calibrator. Then u can run QC. Although u can run qc same time but it would be better to do it later.
Why MY analyzer bubbles daily
how to overcome "the blank of cuvette exceeds limit"
error
Check lamp. When it was installed. Mostly such kind of issue can be fixed by replacing halogen lamp.
12V20W
yeah it was installed before 2 yrs ago, but when I check the halogen lamp it is working well, its continuty works @@biomedicalengineeringbysha5215
@abrihamabate5490
Kindly replace lamp. If u wanna see actual performance, check in debug software.
Quality control liquid and calibration liquid coming allow with this machine. Is it sir ?
Yes it is.
Usually you will get powdered form. You have to dilute with Deionised water.
@@biomedicalengineeringbysha5215 thanks sir
how to send data result to LIS?
Once u will connect analyser with LIS by interfacing after following protocol entered in manual, it will transmit data
Sir could you explain what is reaction cycle and black cycle?
Sir plz make a video on Mindray semi auto chemistry analyzer ...
Plz sir ,,,
Actually Mindray or any brand Semi Automated Chemistry Analyzer is not allowed in GCC countries.
That’s why I can’t but still i will try if i will get a chance.
@@biomedicalengineeringbysha5215 Ok thank u sir
Hello sir apki education qualification kya hai ?
Janaab
Kya koi baat ghalat keh di maine video main? 😂
Na2do4.......dodecyl trimithylammomium chloride.....glysrol
Na2so4....
Hi sir
I m priper my own solution ....and best result .....can i ser with u