Adsorption Isotherm Test

Поділитися
Вставка
  • Опубліковано 25 гру 2024
  • Adsorption Isotherm Test

КОМЕНТАРІ •

  • @jagritijha8210
    @jagritijha8210 2 роки тому +2

    Hi! I am interested to do kinetic studies but have very little idea about it. Could you please guide me to how to actually use it and interpret for my project which is absorbing antibiotic

  • @fareskrika8011
    @fareskrika8011 Рік тому

    How do we get spherical grains that were 400 micrometers in diameter

  • @umehchisom382
    @umehchisom382 Рік тому +1

    Please when you want to adjust for pH. Do you adjust the pH of the dye solution before adding the Adsorbent or when you add the Adsorbent to the dye solution before adjustment for pH

    • @water_waste_labs_media_hub
      @water_waste_labs_media_hub  Рік тому

      It is not a necessary step to adjust pH for adsorption test, except if you know that your adsorbent only works in a certain pH range. If that is the case, I think you should adjust the pH of the solution before doing adsorption. Of course it will change the color of most of the dye solution.

  • @amalinanaimah9266
    @amalinanaimah9266 2 роки тому +1

    Hi, may I know how you calculate adsorption capacity (q) if we took the sample for UV-vis for every 15 mins. Does the V to calculate q change to? Or you just took the initial V of the sample?

    • @water_waste_labs_media_hub
      @water_waste_labs_media_hub  2 роки тому +2

      For an estimation, you can use the initial V to calculate q. Of cause you can take into account the parts that were sampled out, but the calculations would be very complicate.

  • @SerzhKim
    @SerzhKim Рік тому

    at 6:48 "Clean the surface of ..." How it was named? Clean the surface of what?

  • @EdwinMedina147
    @EdwinMedina147 4 роки тому +1

    Could you teach me what is the interpretation of values K and n in freundlich isoterma?

    • @water_waste_labs_media_hub
      @water_waste_labs_media_hub  4 роки тому +1

      K is adsorption capacity(L/mg) and 1/n is adsorption intensity.
      For details, please refer to the following paper.
      www.hindawi.com/journals/jchem/2017/3039817/

  • @saurabhshukla6048
    @saurabhshukla6048 2 роки тому +1

    What was the wavelength at which you took the absorbance readings?

    • @water_waste_labs_media_hub
      @water_waste_labs_media_hub  2 роки тому

      It will depends on which sorbate you used. You can perform a full spectral scan of the target sorbate to determine it's peak absorption wavelength. For Methylene Blue, the wavelength to be used is at 660 nm.

    • @saurabhshukla6048
      @saurabhshukla6048 2 роки тому

      @@water_waste_labs_media_hub Thank you very much for your response. Can you share the wavelength range for Malachite green and Congo Red dyes?

    • @water_waste_labs_media_hub
      @water_waste_labs_media_hub  2 роки тому +1

      Malachite Green (λ max 617 nm)
      Congo Red (λ max 498 nm)

    • @saurabhshukla6048
      @saurabhshukla6048 2 роки тому

      @@water_waste_labs_media_hub Thank you very much for your support.

  • @حمزهفليفل-ص3ه
    @حمزهفليفل-ص3ه Рік тому

    Why did you add water to the standard sample? Because you put the solution on the doses without water, it was not supposed to put water for the standard sample.
    Plase i want understand why add water for strandard sampe.
    Also I have anther Q
    What is the purpose of preparing different concentrations when you did not use them in the experiment?

    • @water_waste_labs_media_hub
      @water_waste_labs_media_hub  Рік тому +1

      Do you mean the one with 0 g of AC? That one is used as baseline reference through the reaction process without any AC. The different conc. solutions are for preparation of calibration curve.