Hematology: Performing and Interpreting QC on the Sysmex XN-550 Automated CBC Analyzer

Поділитися
Вставка
  • Опубліковано 20 січ 2025

КОМЕНТАРІ • 16

  • @jocelynrodriguez3796
    @jocelynrodriguez3796 Рік тому +1

    How to eliminate the point that is outside the graph. Where to write comments? 0:22

  • @khmy1699
    @khmy1699 3 роки тому +1

    Thank you this video is so helpful!💐

  • @richmondafriyie3834
    @richmondafriyie3834 2 роки тому +2

    What do you do when QC fails? What’s the troubleshoot sequence?

    • @basetsanamarakalla2238
      @basetsanamarakalla2238 2 роки тому +3

      1.After shut down check the background check if is ohk and all cell read 0 then proceed with qc.
      2.If qc failed due to any error that occurred during the qc analysis correct the error and rerun the failed level.
      3.check which analytes failed and levels as well... Let's say it the mchc or plt or mch den. The mixing was either to much or too little corrective actions is to remix correctly and rerun.
      3.if level 3 also fails on all analytes check the lot number if it is still the same or it's a new lot number...
      Calibration on this instrument is done by engineers...

    • @richmondafriyie3834
      @richmondafriyie3834 2 роки тому

      @@basetsanamarakalla2238 thanks

  • @nash2454
    @nash2454 4 роки тому +2

    Sometimes sample arms problems with xn550

  • @jocelynrodriguez3796
    @jocelynrodriguez3796 Рік тому +1

    What to do when QC fails?

  • @kudamuyun3173
    @kudamuyun3173 Рік тому

    Love the way you explained. Quite clear.! Are you using Sysmex iQC or 3rd parties iQC as well..Yr LJ chart looks great.

    • @MedicalLabLadyGill
      @MedicalLabLadyGill  Рік тому

      Thank you for the support! Only using Sysmex QC. The institution I worked at before used pools of patient specimens for QC but we don't enough specimens for that practice.

  • @nash2454
    @nash2454 4 роки тому +3

    XN550 always having problems with QC and when plt low we need enter results manually 😑😑

    • @basetsanamarakalla2238
      @basetsanamarakalla2238 2 роки тому

      Check your settings on the instrument and the LIS setup. My instrument transmit a platelet of 1 up to 1200.

  • @anamikatiwari6699
    @anamikatiwari6699 Рік тому

    Manual mode?

  • @jocelynrodriguez3796
    @jocelynrodriguez3796 Рік тому

    What to do when QC fail? 0:22

  • @mike9383
    @mike9383 2 роки тому

    Why cant you use an automated rocker?

    • @basetsanamarakalla2238
      @basetsanamarakalla2238 2 роки тому

      I think the rocker shakes the qc and the manufacturer recommends the we mix vigorously without shakiness... Shaking breaks down the cell and will lead to inaccurate results.

  • @X3technicien04Tech04
    @X3technicien04Tech04 Рік тому

    This method to analyse CQ not correct😮😮