1.After shut down check the background check if is ohk and all cell read 0 then proceed with qc. 2.If qc failed due to any error that occurred during the qc analysis correct the error and rerun the failed level. 3.check which analytes failed and levels as well... Let's say it the mchc or plt or mch den. The mixing was either to much or too little corrective actions is to remix correctly and rerun. 3.if level 3 also fails on all analytes check the lot number if it is still the same or it's a new lot number... Calibration on this instrument is done by engineers...
Thank you for the support! Only using Sysmex QC. The institution I worked at before used pools of patient specimens for QC but we don't enough specimens for that practice.
I think the rocker shakes the qc and the manufacturer recommends the we mix vigorously without shakiness... Shaking breaks down the cell and will lead to inaccurate results.
How to eliminate the point that is outside the graph. Where to write comments? 0:22
Thank you this video is so helpful!💐
What do you do when QC fails? What’s the troubleshoot sequence?
1.After shut down check the background check if is ohk and all cell read 0 then proceed with qc.
2.If qc failed due to any error that occurred during the qc analysis correct the error and rerun the failed level.
3.check which analytes failed and levels as well... Let's say it the mchc or plt or mch den. The mixing was either to much or too little corrective actions is to remix correctly and rerun.
3.if level 3 also fails on all analytes check the lot number if it is still the same or it's a new lot number...
Calibration on this instrument is done by engineers...
@@basetsanamarakalla2238 thanks
Sometimes sample arms problems with xn550
What to do when QC fails?
Love the way you explained. Quite clear.! Are you using Sysmex iQC or 3rd parties iQC as well..Yr LJ chart looks great.
Thank you for the support! Only using Sysmex QC. The institution I worked at before used pools of patient specimens for QC but we don't enough specimens for that practice.
XN550 always having problems with QC and when plt low we need enter results manually 😑😑
Check your settings on the instrument and the LIS setup. My instrument transmit a platelet of 1 up to 1200.
Manual mode?
What to do when QC fail? 0:22
Why cant you use an automated rocker?
I think the rocker shakes the qc and the manufacturer recommends the we mix vigorously without shakiness... Shaking breaks down the cell and will lead to inaccurate results.
This method to analyse CQ not correct😮😮