It cannot be considered as valid but need more information to understand that at what time it drop, for how many seconds or minutes, what was total time of sterilization and at what location it went low and at how many locations it went low
Sir in vessel line there is formation of condensed due to which temperature not reaching 121 so why this condenset is forming in lines what is the reason?
I will make video on this Can you share vessel size , steam trap is there or not, vessel line diameter, length of line, how sensor is placed? Give as much of information available with you as you can
In SIP, usual practice is to set 1 bar pressure and 0.1 bar reduction is allowed after 30 minutes of hold. Unlike autoclave, in SIP, pressure is maintained after SIP throughout before filtration is started hence there is no ingress of non-sterile air in vessel.
Nice sir
Very informative sir
Legend ❤❤❤
Sir filter integrity ke bare me video banayiye na plz
mfg CPP ka video bhejo
Difference between lag time and equilibrium time
Diff. Btwn lag time and equlabrium time
Sir,Sip ka Impact/Risk assessment ka video.
Sir psg ka pressure kitna rhna chahiye
During operation When temperature drop on 121 so in case sip cycle valid or not
It cannot be considered as valid but need more information to understand that at what time it drop, for how many seconds or minutes, what was total time of sterilization and at what location it went low and at how many locations it went low
Sir, Gas filter or J tube ka kya use hota hai storage vessel me
Gas filter for filtering gas when solution level comes down
J tube is to transfer solution from mixing vessel to filtration vessel
Sir in vessel line there is formation of condensed due to which temperature not reaching 121 so why this condenset is forming in lines what is the reason?
I will make video on this
Can you share vessel size , steam trap is there or not, vessel line diameter, length of line, how sensor is placed? Give as much of information available with you as you can
500L vessel yes steam trap is there off inch diameter sensor is placed horizontal to the line
Line length is not sure may be 3/4 feets
Sir what is the leak rate value in sip ... In our company the acceptance limit 0.20 bar that is higher then autoclave why ??
In SIP, usual practice is to set 1 bar pressure and 0.1 bar reduction is allowed after 30 minutes of hold. Unlike autoclave, in SIP, pressure is maintained after SIP throughout before filtration is started hence there is no ingress of non-sterile air in vessel.
Thank you sir