  • @deepakjaiswal3988
    @deepakjaiswal3988 6 місяців тому +1

    Nice sir

  • @mahmmodali2516
    @mahmmodali2516 6 місяців тому +1

    Very informative sir

  • @Monit_Panwar
    @Monit_Panwar 6 місяців тому +1

    Legend ❤❤❤

  • @yogeshkoli851
    @yogeshkoli851 2 місяці тому

    Sir filter integrity ke bare me video banayiye na plz

  • @rakeshbariya-v7z
    @rakeshbariya-v7z Місяць тому

    mfg CPP ka video bhejo

  • @Vikayadavbpharmhhhh
    @Vikayadavbpharmhhhh 6 місяців тому +2

    Difference between lag time and equilibrium time

  • @ravaljinang8389
    @ravaljinang8389 6 місяців тому +1

    Diff. Btwn lag time and equlabrium time

  • @sumantakumarjena484
    @sumantakumarjena484 4 місяці тому

    Sir,Sip ka Impact/Risk assessment ka video.

  • @domendrasahu4058
    @domendrasahu4058 Місяць тому

    Sir psg ka pressure kitna rhna chahiye

  • @namratajain415
    @namratajain415 2 місяці тому +1

    During operation When temperature drop on 121 so in case sip cycle valid or not

    • @PHARMAVEN
      @PHARMAVEN  2 місяці тому

      It cannot be considered as valid but need more information to understand that at what time it drop, for how many seconds or minutes, what was total time of sterilization and at what location it went low and at how many locations it went low

  • @mayankdhiroliya
    @mayankdhiroliya 6 місяців тому +1

    Sir, Gas filter or J tube ka kya use hota hai storage vessel me

    • @PHARMAVEN
      @PHARMAVEN  6 місяців тому

      Gas filter for filtering gas when solution level comes down
      J tube is to transfer solution from mixing vessel to filtration vessel

  • @Rakesh-wj9vn
    @Rakesh-wj9vn 5 місяців тому +1

    Sir in vessel line there is formation of condensed due to which temperature not reaching 121 so why this condenset is forming in lines what is the reason?

    • @PHARMAVEN
      @PHARMAVEN  5 місяців тому

      I will make video on this
      Can you share vessel size , steam trap is there or not, vessel line diameter, length of line, how sensor is placed? Give as much of information available with you as you can

    • @Rakesh-wj9vn
      @Rakesh-wj9vn 4 місяці тому

      500L vessel yes steam trap is there off inch diameter sensor is placed horizontal to the line
      Line length is not sure may be 3/4 feets

  • @mohitbisht2041
    @mohitbisht2041 4 місяці тому

    Sir what is the leak rate value in sip ... In our company the acceptance limit 0.20 bar that is higher then autoclave why ??

    • @PHARMAVEN
      @PHARMAVEN  4 місяці тому +1

      In SIP, usual practice is to set 1 bar pressure and 0.1 bar reduction is allowed after 30 minutes of hold. Unlike autoclave, in SIP, pressure is maintained after SIP throughout before filtration is started hence there is no ingress of non-sterile air in vessel.

    • @mohitbisht2041
      @mohitbisht2041 3 місяці тому

      Thank you sir