Brabender MetaStation 4E: Usage 1 | How to prepare a test

Поділитися
Вставка
  • Опубліковано 19 вер 2024
  • Learn more: www.brabender....
    Measure the rheological properties of PVC materials with the MetaStation 4E and the Measuring Mixer W 50 EHT.
    How to prepare a test:
    - Insert the mounting surface for the measuring mixer
    - Fix the mounting surface with a star grip nut
    - Push the coupling sleeve onto the clutch of the MetaStation
    - Attach the measuring mixer
    - Tighten the knurled nuts on the stud bolts
    - Connect the cable of the 2-hand safety device of the measuring mixer with the socket "Safety Device" on the control side of the MetaStation
    - Connect the cable of the torque limitation to the socket "Torque limit key"
    - Connect the cable of the safety device to the correspondent socket
    - Mount the control and stock thermocouples into the corresponding bayonet nipples of the measuring mixer
    - Connect the cables of the control thermocouples to the corresponding sockets "Control Temperature" at the terminal board of the Meta-Station
    - Connect the heating cables
    - Connect the cables of the stock thermocouples to the corresponding sockets "Stock Temperature"
    - Connect the cooling air hoses to the correspondent sockets
    - Make sure that the stock temperature thermocouple reaches approx. 1.0 to 1.5 mm into the mixer bowl
    - Attach the mixer support and tighten the screws
    - Loosen the screws at the front plate and draw it off
    - Draw off the mixer bowl
    - If not yet mounted, insert the longer blade into the left bushing and the shorter one into the right bushing
    - Mount the mixer bowl and the front plate again