ODME-S3000 (KSB). Calibration Test

Поділитися
Вставка
  • Опубліковано 17 лис 2024

КОМЕНТАРІ • 5

  • @mohamedkhames2602
    @mohamedkhames2602 6 місяців тому

    Welcome and thank you for the explanation of calibrating the device. I have a new device and I installed it, but when calibrating it, calibration fail appears in the calibration. Are there steps to adjust the sensors?

    • @kyawyenyein3172
      @kyawyenyein3172  6 місяців тому

      There are a couple of reasons in calibration failure. Mostly 1.Wiper deteriorating/ replace with new one or if fall down/need to open and fix it back. 2. Projector light dim /to replace with new one. 3. Check Pneumatic Valve positions 4. Check the connections between cell block and Oil content meter. In my case, Optical fiber sensors in oil content meter are found dirty with small particles which cause the calibration failure.

  • @azereduw1
    @azereduw1 Рік тому

    Do you have a step-by-step guide on how to do this?

  • @gabrielazeredocutalosouza2829

    Do you have a step-by-step guide on how to do this?

    • @kyawyenyein3172
      @kyawyenyein3172  11 місяців тому

      That's how we do step by step bro.
      Do you mean setting up the lines on the deck where ODME unit installed?