How to run EIS analysis for solid or film sample using Gamry Reference600 potentiostat

Поділитися
Вставка
  • Опубліковано 24 жов 2024

КОМЕНТАРІ • 55

  • @miricagroup6809
    @miricagroup6809 8 днів тому +1

    Thank you for demonstrating the experimental procedure. It is very useful for my experiment. I would like to learn more about:
    1) how should we pick the frequency range and ac voltage for the experiment?
    2) which model should we use for the data fitting?
    Your help is highly appreciated!

    • @mohdhazwan6523
      @mohdhazwan6523  7 днів тому +1

      @@miricagroup6809
      Hi,
      1. Yes, this is crucial to know what is actually happening in the inner Helmholtz or outer Helmholtz plane. Different sample will have different frequency range of interest.
      2. For the circuit model, this is tricky since it is all depends on the type of sample as well as the Bode plot pattern that you will get. If you didn't have any idea on whixh circuit to be used, we can guess the electrical component based on the pattern of your Bode and Nyquist plots. Probably you can check my recorded videos (KAA504) on EIS for more information

  • @anirudhsubramaniam3740
    @anirudhsubramaniam3740 3 місяці тому +2

    Thank you for uploading Dr. Hussin! Helped me a lot with my experiment!

  • @21HeartSong
    @21HeartSong 9 місяців тому +2

    Thank you very much for an amazing video! It was very useful and well explained.

  • @shadihosseini2066
    @shadihosseini2066 9 місяців тому +2

    amazing video. it was reallllyy helpful

  • @davidadekoya3468
    @davidadekoya3468 Рік тому +2

    Thank you very much for this video.

  • @beyadgalemendawoke6846
    @beyadgalemendawoke6846 9 місяців тому +2

    Thanks dear, what do you say about t at different temperature intervals, 20, 30 , 40 , 50.......80?

    • @mohdhazwan6523
      @mohdhazwan6523  9 місяців тому

      You need a stainless steel coin cell that can be put in an oven. From there you can study the thermodynamics

  • @ashuntu2010
    @ashuntu2010 3 місяці тому +1

    Thanks for very informative video. May I know from where to get such type of cell? Thanks.

    • @mohdhazwan6523
      @mohdhazwan6523  3 місяці тому

      @@ashuntu2010 this is a fabricated cell..you can use a normal coin cell to run this experiment

  • @matheuschristopher7181
    @matheuschristopher7181 Рік тому +2

    Dear Professor Hussin, your video is very helpful. May I ask some references that explain the Electrochemical Impedance Spectroscopy mechanism with this tool and and parameter values ​​that need to be inputed in the software?
    Thank you very much ✨

    • @mohdhazwan6523
      @mohdhazwan6523  Рік тому

      Hi, please check my publications related to polymer electrolyte studies. I have used similar setup and all parameters are written in the article. All the best to you.

  • @safiahabdurhamnalshehri268
    @safiahabdurhamnalshehri268 Рік тому +1

    Thank you very much Professor Hussin for this useful video.
    I'd appreciate it if you explain how to save the fitting data in Gamry. + What are the components of "CPE with diffusion" circuit exactly? would it be possible to share with us the title of your publication in this regard?
    Many thanks

    • @mohdhazwan6523
      @mohdhazwan6523  Рік тому +1

      Thank you for your post. Well, to save the fitted data you can just save as the file after the perfect fitting has been done. It will be saved under .GData format

  • @BerniceAbraham12
    @BerniceAbraham12 4 місяці тому +1

    Thank you very much for this video, it is really helpful. When the Nyquist plot is a semi-circle, can the CPE with diffusion model be used to fit the curve, or else which model is preferable?

    • @mohdhazwan6523
      @mohdhazwan6523  4 місяці тому

      If the Nyquist plot shows a semicircle with a 45° line at the end (due to diffusion) of the plot, it is best to use CPE+Warburg impedance circuit for better data fitting

    • @BerniceAbraham12
      @BerniceAbraham12 4 місяці тому +1

      @@mohdhazwan6523 Thanks for your response. Can this model be applied if it's just the semi-circle without the line at a 45⁰ angle?

    • @mohdhazwan6523
      @mohdhazwan6523  3 місяці тому

      @@BerniceAbraham12 nope..better to fit it with normal Randles CPE circuit..without Warburg resistance

    • @BerniceAbraham12
      @BerniceAbraham12 3 місяці тому

      Thanks Sir, for taking out the time to reply.

  • @riccardogherardini2960
    @riccardogherardini2960 Рік тому +1

    Dear Hussin, thanks for the amazing video !! I have doped CeO2 powder, do I have to produce a film or a pellet to do the experiment??

    • @mohdhazwan6523
      @mohdhazwan6523  Рік тому +1

      In this case i would recommend you to make it into a pellet. Probably you can you FTIR pellet mould to do that

  • @mehrpay
    @mehrpay Рік тому +1

    Professor Hussin, thanks for sharing your knowledge with this video, I'd appreciate it if you could tell why there is a relatively large value of Error with exponenstial of 18 for Rp in the video. Is this normal?

    • @mohdhazwan6523
      @mohdhazwan6523  Рік тому +2

      Hi, I would normally ignore the error values but will always refer to to the goodness of fit. If the goodness of fit is relatively low, then it means that the equivalent circuit used has a better fitting

  • @ahmadfathaziznugraha2581
    @ahmadfathaziznugraha2581 Рік тому +1

    Dear Professor, thank you for sharing your knowledge. I wanted to know does this cell setup works for other type of potentiostat?

  • @VaryingpHValues
    @VaryingpHValues 7 місяців тому +1

    Thanks for your video. By the way, where did you get this kind of cell for EIS test? I may purchase online. My sample is like sticky gel and I think I can also use similar ones. Another question is about the resistance of the cell itself. Had you deducted this value when you calculated the resistance of your sample?

    • @mohdhazwan6523
      @mohdhazwan6523  7 місяців тому

      That's a good question. Yeah, we have checked the resistance of the cell holder itself and because of the resistance value is too low the actual resistance of the sample is not differ much

    • @VaryingpHValues
      @VaryingpHValues 7 місяців тому

      OK. Many thanks. Did you purchase this cell? I may need a similar one to conduct my experiment.@@mohdhazwan6523

  • @XSD388
    @XSD388 3 місяці тому +1

    Hello Dr. Hussin just to confirm you are using a 2 probe setup with working and counter correct ?

    • @mohdhazwan6523
      @mohdhazwan6523  3 місяці тому +1

      @@XSD388 yes two electrode system..WE-(CE+RE)..meaning WE and CE short circuited with RE

  • @akshaydixit5115
    @akshaydixit5115 Рік тому +1

    Hello Professor Hussin, I wanted to know whether you have conducted EIS with soil environment. If yes, please guide me through the procedure and difficulties associated.

    • @mohdhazwan6523
      @mohdhazwan6523  Рік тому

      Yes, i have tested soil sample as well but under two different conditions.
      1. I run dried soil sample for conductivity measurement using EIS. For this purpose I use two electrode sandwich system.
      2. I did corrosion monitoring of steel sample buried in soil. But it is not dried, it has at least 40% of water and for the reference electrode i have used silver wire, while counter electrode is graphite.

    • @akshaydixit5115
      @akshaydixit5115 Рік тому +2

      @@mohdhazwan6523 If you have taken the picture or created a video of the setup then can you please share it with me.

  • @LUISFRANCISCOSANCHEZSOVERO
    @LUISFRANCISCOSANCHEZSOVERO Рік тому +1

    Dear Hussin, thanks for sharing your knowledge , I watched in the video , 9:40min , before user measured a sample thin film. is p-type silicon as a substrate and the film is Silicon carbide?. In addition is possible to send me the dimensions of the solid/film sample holder.

    • @mohdhazwan6523
      @mohdhazwan6523  Рік тому

      Yes we can run p-type or n-type semiconductor samples using EIS too. The dimension of the film was 1.5 cm x 1.5 cm

  • @sstrust6549
    @sstrust6549 5 місяців тому +1

    Hello, can you tell me where you purchased the membrane sandwich holder...

    • @mohdhazwan6523
      @mohdhazwan6523  5 місяців тому +1

      It is a fabricated cell..you can find and vendor to fabricate it

    • @sstrust6549
      @sstrust6549 5 місяців тому

      @@mohdhazwan6523 please can you share the pics and dimensions with me, we are going to make it ..

  • @okanyy6367
    @okanyy6367 Рік тому +2

    Hello, can we say that Ru or Rp value is the corrosion resistance of material to solution ?

    • @mohdhazwan6523
      @mohdhazwan6523  Рік тому

      In the case of corrosion, we look at Rp value. But since this video is more on finding conductivity of solid sample, the Rp or Rct is assumed as bulk resistance to be insert in conductivity formula.

  • @farzana7620
    @farzana7620 5 місяців тому +1

    Assalamualaikum Dr, i use the same setup as your for the solid polymer electrolyte usinf HZ7000 potentiostat. But for the nyquist plot my data is not a semicircle/straight it was scattered instead. May i know the possibility of this matter come from whether the sample of SPES itself or else. I use the same parameter as stated in the video. Thank you Dr

    • @mohdhazwan6523
      @mohdhazwan6523  5 місяців тому

      Well, scattered data points could from various reasons like poor interfacial contact between the electrode and the membrane, connection problem, sample is too thick or not homogeneous, disturbance from surrounding, wrong frequency range etc..perhaps you can check some of this reason

    • @farzana7620
      @farzana7620 5 місяців тому +1

      @@mohdhazwan6523 Thank you Dr. really appreciate your explanation.

  • @melkiegetnet5492
    @melkiegetnet5492 Рік тому +1

    Hey, could you mind if you can send me the dimensions of the solid/film sample holder?
    Thanks and looking forward hearing from you

    • @mohdhazwan6523
      @mohdhazwan6523  Рік тому +1

      This is a fabricated sample holder. I am afraid I didnt kept the drawing. But you also purchase a commercial coin cell for the same purpose.

  • @beshoynasr
    @beshoynasr Місяць тому +1

    Hi please i want the setup for measuring impedence for membrane film where i can purchase in Japan

    • @mohdhazwan6523
      @mohdhazwan6523  Місяць тому

      @@beshoynasr not sure where to buy in Japan but you can always search for stainless steel coin cell

    • @beshoynasr
      @beshoynasr Місяць тому

      @@mohdhazwan6523 thanks for responding I already use stalesteel one but there is no connection between stalesteel and nafion

  • @fadhilahrahmat515
    @fadhilahrahmat515 Рік тому +1

    Assalamualaikum Dr Hazwan. Saya ingin bertanya, jika sampel saya adala film (starch+ graphene nanoplatelet) boleh tak saya gunakan kaedah rendaman (with an electrolyte and 3 electrodes) dalam melakukan ujian EIS

    • @mohdhazwan6523
      @mohdhazwan6523  Рік тому +1

      Wasalam, kalau nak lihat dielectric properties utk film tersebut lebih baik buat dalam keadaan solid state. Ini kerana bila kita rendam dalam electrolyte solution, polymer matrix drp starch serta ionic transport utk film tersebut sudah berbeza (dalam kes ini electrolyte yg memainkan peranan utk ionic transport, bukan film tersebut).