Unboxing, calibration and testing pH meter - smart sensor 818

Поділитися
Вставка
  • Опубліковано 31 січ 2022
  • In this video we unbox the Smart Sensor 818 Ph-meter, demonstrate calibration and then do an accuracy test.
    Link to our pH meters (in Norwegian):
    www.innehagen.no/ph-og-ec
  • Наука та технологія

КОМЕНТАРІ • 58

  • @allann2099
    @allann2099 2 роки тому +1

    Merci, une bonne vidéo pour faire ma calibration. Claire et précise. Bravo👍👍

    • @innehagen4248
      @innehagen4248  2 роки тому +1

      Merci beaucoup! Nous avons été ravis d'apprendre que vous trouviez la vidéo utile. (We hope this is correct french, used google translate:))

  • @GRJCLyon
    @GRJCLyon Рік тому +1

    This must be one of those ASMR videos. No talking only sounds. Thanks a bunch.

    • @innehagen4248
      @innehagen4248  Рік тому

      Thank you. We tried to make it as self explanatory as possible and we dont like constant babbling :-) We are glad you found it relaxing.

  • @iumi2329
    @iumi2329 Рік тому

    Thank you.

  • @bioripososrl7355
    @bioripososrl7355 Рік тому +1

    molto interessante.

  • @NRocky94
    @NRocky94 Рік тому

    Is it still working a months later ? Did you have to replace the probe already ? Pretty curious as I'm considering buying one of those.
    Very nice video btw. Very relaxing.

    • @innehagen4248
      @innehagen4248  Рік тому +1

      Thank you very much🙂 Yes, we are still using this pH-meter every week. We have not replaced the probe, but we have been careful to have enough pH storage solution in the cap and we have calibrated the meter a couple of times 😀 If you look at this video we use the same pH meter 1,5 years later m.ua-cam.com/video/5z3L67xs3-U/v-deo.html

  • @BohumirZamecnik
    @BohumirZamecnik Рік тому

    Nice! How much off can it be before calibration. Mine came without the bags. I'll have to get them separately.

    • @innehagen4248
      @innehagen4248  Рік тому

      We are not sure how much off it was before we calibrated it. It should be pretty accurate from the factory but it is recommend to calibrate it to be sure 😀

  • @BallisticTeror
    @BallisticTeror 2 роки тому +1

    the ph meter your using looks exactly like my DR meter ph838

    • @innehagen4248
      @innehagen4248  2 роки тому

      Yes, it looks the same :-) I think the difference is the accuracy 0,01 vs 0,05.

  • @user-vr7nn9vx8x
    @user-vr7nn9vx8x 9 місяців тому +1

    Thank you so much for the video. My 6.86 buffer powder was not in powdered form when i opened it and now its saying error, what coukd be the issue?

    • @innehagen4248
      @innehagen4248  9 місяців тому +1

      Thank you 😀 Can it be that the bag was damaged? And that the humidity got into the bag? Also did you read our other comment about using demineralised water? It is possible to buy new bags cheap on aliexpress.

    • @user-vr7nn9vx8x
      @user-vr7nn9vx8x 9 місяців тому +1

      I used a deionized water

  • @feliciaangelique5948
    @feliciaangelique5948 Місяць тому +1

    I want to ask something
    The temperature of the environment that displayed on my ph meter is 47 degree celcius which is not like the actual temp
    And i try to calibrate it and always shown error
    Is it can because of the damage of my ph meter or what?
    Thanks before

    • @innehagen4248
      @innehagen4248  Місяць тому

      It sounds like there might be a problem with your ph meter. You can try refilling ph refill/storage solution in the cap (3 mol KCl). If this doesnt help you should contact the store where you bought it.

  • @mixdup0016
    @mixdup0016 Рік тому

    Should it come with storage solution inside the cap mine is dry with something yellow at the bottom

    • @innehagen4248
      @innehagen4248  Рік тому

      The yellow at the bottom is a sponge. The sponge should be wet with storage solution when it comes from the factory. You should not let the sponge dry out, you should refill with storage solution once in a while. If the sponge was hard/dry when you got it, that would be a bad sign. Try to add storage solution and let it rest for a day. This can fix the probe. If not you should contact the store where you bought it.

  • @jamesbishop5119
    @jamesbishop5119 2 роки тому

    The cap on mine came with some yellow crystal looking stuff on the bottom, I am unsure of how to store electrode

    • @jamesbishop5119
      @jamesbishop5119 2 роки тому

      It does not look like a sponge at all and is dry...

    • @innehagen4248
      @innehagen4248  2 роки тому

      That does not sound good. If it was crystalized it has probably dried out. It might have been stored too long in the store you bought it...or the rubberring which seal the cap is damaged. There should be a little bit storage solution in the cap like this www.innehagen.no/product-page/growth-technology-ph-probe-refill-storage-300-ml

  • @sharnazklein1587
    @sharnazklein1587 9 місяців тому +1

    My pH meter is reading err
    What could be the problem? This is after pressing the cal button and putting it in the first solution

    • @innehagen4248
      @innehagen4248  9 місяців тому

      Are you using demineralised water or just tap water? Demineralised water is water completely free of dissolved minerals as a result of, distillation, deionization or membrane filtration (reverse osmosis). Old car batteries use demineralised water so your local hardware store might have it if you ask for battery water. In this movie we use "battery water". You can also just try to google demineralised water to see where you can buy it nearby. If you use normal tap water, it will have the wrong pH and your calibration liquid will be out of range for calibration.
      It is also important to be very accurate when you mix the powder with the water. You have to get all of the powder out of the bag. If there is just a little left in the bag the mixture will have the wrong pH.
      Also you must use exactly the right amount of water (250ml in our case).

  • @kidtuber3407
    @kidtuber3407 Рік тому +1

    Does the storage solution come with the product?

    • @innehagen4248
      @innehagen4248  Рік тому

      Hi, there is a little bit in the cap when you buy it :-), but with regular use this will dry out after a while (couple of months depending on use). Therefore we recommend to buy some with the meter.

    • @ngimtaing3347
      @ngimtaing3347 Рік тому

      Hi Do you have the website to buy the solution for testing the HP meter

    • @BohumirZamecnik
      @BohumirZamecnik Рік тому

      It was jn the product description, rhe manual, but the powder bags were actually missing. 😢

  • @bigtexas1376
    @bigtexas1376 Рік тому +1

    Does this have to be that water or can it be bottles water ?

    • @innehagen4248
      @innehagen4248  Рік тому

      Hi, it has to be demineralized water completely free of dissolved minerals. We bought it at local hardware store, in our case it was intended to refill car batteries 😀 But we guess you can get it anywhere. We are sure you find it nearby if you search for demineralized water👍

    • @bigtexas1376
      @bigtexas1376 Рік тому

      Ok thankyou very much 💯

    • @Krowmobe
      @Krowmobe Рік тому

      Pure R.O. water works as well or distilled with no minerals added back to it.

  • @ayodeleawolola7050
    @ayodeleawolola7050 8 місяців тому

    Hi, good morning. Whats the frequency of calibration of the meter, pls? After each use?

    • @innehagen4248
      @innehagen4248  8 місяців тому +1

      Hi, you do not have to calibrate it after every use, 3-4 times a year should be enough or you can calibrate it again if it has been stored for a long time 🙂

    • @ayodeleawolola7050
      @ayodeleawolola7050 8 місяців тому +1

      @innehagen4248 thank you so much. This has been helpful.

  • @kmslide5796
    @kmslide5796 Рік тому

    When i do it and dip it into the 4.00 its just blinking ans the pas is Not coming

    • @innehagen4248
      @innehagen4248  Рік тому

      Hm...are you using deminerlized water? Measure exactly 250ml of water and all of the powder from the bag?

  • @vincentlee2460
    @vincentlee2460 2 роки тому

    Till now is it still accurate?

    • @innehagen4248
      @innehagen4248  2 роки тому

      Yes, it is still working as it should 😀 We always make sure there is storage solution in the cap!

  • @generalawareness101
    @generalawareness101 Рік тому +3

    What I so dislike about pH meters is that they lose accuracy and in about 18 months need to be replaced, or the bulb probe does. VERY expensive to have to keep replacing them like that.

    • @innehagen4248
      @innehagen4248  Рік тому +1

      We have been using this pH meter for nearly a year now and so far so good. Keeping our fingers crossed :) I think the main problem with many pH meters that stops working is that pH meters are not sold together with pH storage solution, and people are not generally aware ;) that the probe will dry out and stop working if it is not stored with pH storage solution in the cap. We regulary check and refill storage solution in the cap so I have hopes that it will last longer than 18 months. See 3:00 . Example of the storage solution we use: www.innehagen.no/product-page/growth-technology-ph-probe-refill-storage-300-ml

    • @generalawareness101
      @generalawareness101 Рік тому

      @@innehagen4248 Yes, and no. This was brand new just opened, and from Blue manufacturer if kept dry it loses accuracy but stick it in the solution for 10 mins, and it will be back.

    • @LunaNL
      @LunaNL Рік тому

      18 months replacement your lucky my friend 2 months over here i have a hole colection of ph meters

    • @AmberlinaM
      @AmberlinaM Рік тому

      What is the ‘storage solution’??

    • @generalawareness101
      @generalawareness101 Рік тому +2

      @@LunaNL Yeah, 18 months according to them in the real world I suspect half that.

  • @franciscodiaz4455
    @franciscodiaz4455 3 місяці тому

    Este phmetro no trae protector de electrodo gracias

    • @innehagen4248
      @innehagen4248  3 місяці тому

      The pH meter in this video does come with electrode protection. The cap has a wet sponge with storage solution and it is is sealed with a rubber o-ring. If your pH meter does not have electrode protection it will not be accurate.

    • @franciscodiaz4455
      @franciscodiaz4455 3 місяці тому +1

      El agua desmineralizada para bateria de auto sirve para diluir la solución de calibración gracias

    • @innehagen4248
      @innehagen4248  3 місяці тому

      Yes, it is demineralized water for car batteries👍/ Sí, es agua desmineralizada para baterías de auto👍

    • @franciscodiaz4455
      @franciscodiaz4455 3 місяці тому

      @@innehagen4248 gracias

  • @errthinpurplee2352
    @errthinpurplee2352 2 роки тому

    mine says err ., why

    • @innehagen4248
      @innehagen4248  2 роки тому

      It might say err if the calibration liquid is outside range. Did you mix the calibration powder in deionized water?

    • @errthinpurplee2352
      @errthinpurplee2352 2 роки тому +1

      @@innehagen4248 i am sorry im so dumb i had the ph meter on the wrong cup of ph ., i got it done now its working fine ., thanks for replying

    • @Lafavy
      @Lafavy 9 місяців тому

      El mío sigue indicando err😢

    • @innehagen4248
      @innehagen4248  9 місяців тому

      Are you using demineralised water or just tap water? Demineralised water is water completely free of dissolved minerals as a result of, distillation, deionization or membrane filtration (reverse osmosis). Old car batteries use demineralised water so your local hardware store might have it if you ask for battery water. In this movie we use "battery water". You can also just try to google demineralised water to see where you can buy it nearby. If you use normal tap water, it will have the wrong pH and your calibration liquid will be out of range for calibration.
      It is also important to be very accurate when you mix the powder with the water. You have to get all of the powder out of the bag. If there is just a little left in the bag the mixture will have the wrong pH.
      Also you must use exactly the right amount of water (250ml in our case).