    Fill out this form and I will help you using my network of Hospitals and Companies:
    docs.google.com/forms/d/e/1FA... Most electrical safety measurements can be done with a regular multimeter (like ground resistance and Attached Ground Leakage). To really see what's going on with a medical device you need to lift the ground and test the device with the power on. For this, I have a jig that's easy to make and easy to use.
    Many developing countries and remote locations might not have the budget or logistics to purchase and maintain a calibrated official electrical safety analyzer. Yet, all electrical repair centers have some sort of multimeter.
    With this jig, those locations can now preform a normal electrical safety. ** PLEASE SUBSCRIBE! I am trying to grow this channel to deliver better content. Subscribers = Corporate Supporters =Better Content
    ** Check out my playlists for Biomed Basics, Vlog, Equipment Repair, and HTM Theory.
    ** Connect with me on LinkedIn to stay updated:
    / justin-barbour-8602b886
    **
    ********************************************
    Help support Better Biomed Channel!
    Fuel, Travel Fees, Equipment, Website hosting fees - it all adds up! Every dollar is greatly appreciated.
    betterbiomedchannel.betterworl...
    ******************************************** Check out my other channel @TakeItApartChannel

КОМЕНТАРІ • 31

  • @theworshipdrummern8
    @theworshipdrummern8 4 місяці тому +1

    This tutorial also comes in extremely handy for when I have to send my ESA620 in for cal once a year. I’m a one man biomed shop and when I’m down my analyzer for the better part of a week, this will allow me to still be able to complete PMs or post repair re-certs. Awesome video!

    • @BetterBiomedChannel
      @BetterBiomedChannel  4 місяці тому

      I need to make an updated version of this. There should be a 1000Ohm resistor in line.

  • @StevenAhmodu
    @StevenAhmodu Рік тому +2

    Thank you, Better Biomed. I wish I could give this video a thousand likes. You are good

  • @HaseebElectronics
    @HaseebElectronics 3 роки тому +2

    beautiful and perfect tutorial on electrical safety. your demonstration is fantastic

  • @zainElect01
    @zainElect01 3 роки тому +2

    wow it is brilliant and perfect professional demonstration to make multimeter jig. thanks for sharing your valuable experience

  • @haashirvlogzhaseebelectron7373
    @haashirvlogzhaseebelectron7373 3 роки тому +2

    fabulous and superb work. stay blessed and keep sharing your excellent experience

  • @carlosargumedo5665
    @carlosargumedo5665 3 роки тому +2

    OH OH THANK YOU VERY MUCH ENGINEER, IT WILL BE A GREAT HELP IN THE TEAM

  • @haseebelectronicsurdu
    @haseebelectronicsurdu 3 роки тому +2

    wonderfully awesome and superb demonstration. thanks for sharing. keep it up

  • @janasmeal7916
    @janasmeal7916 3 роки тому +2

    brilliant and superb professional work. thanks for sharing

  • @HTMWorkshop
    @HTMWorkshop 3 роки тому +3

    I have always wondered about using a multimeter. Great Idea! Seems like the only reason you would need a true ESA is for patient leakage tests or ISO testing which requires thing like reverse polarity - but even all of those you could probably create a jig. You have my mind thinking lol.

    • @BetterBiomedChannel
      @BetterBiomedChannel  3 роки тому +5

      About 50% of my viewers are at locations outside of the USA. Some are on islands and some are in very poor countries with no electrical safety codes for medical. - They do seem to all have some sort of multimeter. By showing them ground resistance and a jig for case leakage, they can start checking their equipment today. We really can help change the world.

  • @joeyxbat
    @joeyxbat Рік тому

    Wow, great idea!

  • @angelusmendez5084
    @angelusmendez5084 3 роки тому +1

    Great tutorial, thanks very much! 👏

  • @ak_vg9715
    @ak_vg9715 2 роки тому +1

    Great idea,thanks for sharing.

  • @life2lights
    @life2lights 3 роки тому +2

    Very interesting

  • @BiP00
    @BiP00 13 днів тому

    Man you rock!!

  • @bc7522
    @bc7522 3 роки тому +1

    Just did this last week make shift earth for testing Philips hearsart xl+ defib

  • @MrTakaiko
    @MrTakaiko 3 роки тому +1

    Thats nice idea !! But the issue we still need record !! Reports for cmms for our records to cover our backs lol right !

  • @williambirt914
    @williambirt914 Рік тому +1

    Thank You . Could also get a 4 x 4 box with an outlet one side and a switch (to open and close the ground) with a cord to the wall and plug the electrical device into outlet. switch opens ground. this would work better for those without soldering?
    Thanks

    • @BetterBiomedChannel
      @BetterBiomedChannel  Рік тому

      That's a very cool idea!
      See! Open source worldwide biomed problem solving. A 4x4 gang box and switch would be available anywhere.

  • @diydan6045
    @diydan6045 2 роки тому +1

    Can you talk about electrical safety regulations. And what the hospital needs to follow if it is by state or federal. A lot of our equipment PM status has been changed to no PM required.

    • @BetterBiomedChannel
      @BetterBiomedChannel  2 роки тому

      According to the latest NFPA99 and CMS requirements, many items fall under a possible AEM program. - unfortunately. Some things like light sources and warming cabinets should always be cleaned - because they can catch fire 🔥.... but some places still put "no pm required" on them.....

    • @diydan6045
      @diydan6045 2 роки тому

      @@BetterBiomedChannelThey are literally only going by the manufacturer's recommendations in the service manual.They put the OR Solutions fluid warmer on a 5 year a PM. The manual is talking about the module that keeps the temp not electrical safety. In my book if the medical equipment is portable and also has a removable AC cord it needs to be safety tested. I have seen 1 hospital grade molded power cord with the hot and ground reversed.

  • @MemoriesRR
    @MemoriesRR 3 роки тому

    Don’t you need an AAMI simulated patient in series with the test lead. It is a specific capacitor and resister .

    • @BetterBiomedChannel
      @BetterBiomedChannel  3 роки тому

      No, I've never heard of such a thing unless that is "patient leakage". Case Leakage is a direct measurement of energy present on any electrical device compared to ground potential.

  • @happymeal4406
    @happymeal4406 3 роки тому

    While I do think this is very cool I don't see this as being feasible in the US. I don't think TJC would be very receptive to some kind of bootleg ESA like this. Are ESA's not readily available in other countries? In my organization test equipment must always be documented. I can only imagine the storm if an accreditor discovered something as this was being used.
    Again, I think this is super interesting and informative into the the concept and explanation for electrical safety. I just don't see how this is practical or could ever be used.

    • @BetterBiomedChannel
      @BetterBiomedChannel  3 роки тому

      You're reaching a little bit. Every inspection I have ever been in (which I a lot) doesn't care about the biomed program (for the most part). I have never heard of them inspecting our test equipment. Never. Most work orders at most places I have been at do not even record the safety analyzer used. ESAs are not readily available in most other countries and even harder is annual calibration.

    • @BetterBiomedChannel
      @BetterBiomedChannel  3 роки тому +1

      A real application is with 220V devices. No shop that I've ever heard of has the 220V plug adaptors to test all the 220V devices that enter the building. Most places BS their way through those electronically safeties but I developed 220V jigs of each type of wall adaptor with liftable ground.

    • @Gold_Dust1
      @Gold_Dust1 5 місяців тому

      This was a great video! It would be fantastic if you could share the construction of your 220V Jigs one day! I've been planning on making a jig for my floor centrifuges. @@BetterBiomedChannel