Ref. out of range Reference out of range. The counter of the internal reference capacitor (signal conversion) is not within the expected limits. Parameters wrong. Switch off the instrument, wait one minute then switch the instrument on. Check parameter settings. Call service technician. Check that the offset setting in the temperature menu is the same as in the instructions on the photometer board. Refer to chapter 8.5.2, "TEMPERATURE category settings" to carry out a functional check of the temperature controller and sensor offset set- tings. Replace the photometer board, if necessary.
Flow cell did not reach 37 °C +/- 0.2 °C (offset is taken into account) within 1 hour after start up. Peltier element/sensor defect. Note: a test is made before every measurement to check that the temperature is stable at 37 °C +/- 2°C. If the temperature is out of range, a T-flag is added to the results. Note: when a temperature error occurs the peltier element is switched off to prevent damage to the flow cell.
Switch off the instrument, wait one minute then switch the instrument on. Check that the environ- ment conditions are within the specified limits. Check that the filter at the rear of the instrument is not clogged. Ensure that the environ- ment conditions are met. Call service technician
Check temp sensor. If the sensor is malfunctioning replace the photo- meter board; remember to set the offset in the temperature menu. Check resistance of the peltier element. Replace if necessary. Check connections between the processor board and the peltier ele- ment/sensor (please refer to the Appendix, A.4 "Electrical Drawings"). Check connections to the integrated photometer board are secure. Check for obstructions in the airflow to the heat sink of the peltier element. Check temperature paste between heat sink peltier and house assy. Check that the offset setting in the temperature menu is the same as in the instructions on the photometer board. Refer to chapter 8.5.2, "TEMPERATURE category settings" to carry out a functional check of the temperature controller and sensor offset set- tings. Note: the overheat protection can be switched off in the service menu; overheating can cause damage to the optical system.
@@tech.keys. شكراً لك من قلبي أخي على هذا المجهود الذي فعلته معي وايضاً على تواضعك واخلاقك الرائعة في معاملتك مع من يطلب العلم حقاً انت انسان رائع وجيد
Bhai apna no. Send krdo aapke work se related bt krni hai ya adrress btade kaha rehte hai aap
03160536556
Micro lab 300 error cod 2 ka prablam ka solution sir
Ref. out of range Reference out of range.
The counter of the internal
reference capacitor (signal
conversion) is not within
the expected limits.
Parameters wrong.
Switch off the instrument,
wait one minute then
switch the instrument on.
Check parameter settings.
Call service technician.
Check that the offset setting in the temperature menu is the same as in
the instructions on the photometer board.
Refer to chapter 8.5.2, "TEMPERATURE category settings" to carry out
a functional check of the temperature controller and sensor offset set-
tings.
Replace the photometer board, if necessary.
لوسمحت اخي ماهو سبب هذي المشكلة ERRORCODE:0 لو سمحت أكتبلي الحل في التعليق
Flow cell did not reach
37 °C +/- 0.2 °C (offset is
taken into account) within
1 hour after start up.
Peltier element/sensor
defect.
Note: a test is made before
every measurement to
check that the temperature
is stable at 37 °C +/- 2°C.
If the temperature is out of
range, a T-flag is added to
the results.
Note: when a temperature
error occurs the
peltier element is switched
off to prevent damage to
the flow cell.
Switch off the instrument,
wait one minute then
switch the instrument on.
Check that the environ-
ment conditions are within
the specified limits.
Check that the filter at the
rear of the instrument is
not clogged.
Ensure that the environ-
ment conditions are met.
Call service technician
Check temp sensor. If the sensor is malfunctioning replace the photo-
meter board; remember to set the offset in the temperature menu.
Check resistance of the peltier element. Replace if necessary.
Check connections between the processor board and the peltier ele-
ment/sensor (please refer to the Appendix, A.4 "Electrical Drawings").
Check connections to the integrated photometer board are secure.
Check for obstructions in the airflow to the heat sink of the peltier
element.
Check temperature paste between heat sink peltier and house assy.
Check that the offset setting in the temperature menu is the same as in
the instructions on the photometer board.
Refer to chapter 8.5.2, "TEMPERATURE category settings" to carry out
a functional check of the temperature controller and sensor offset set-
tings.
Note: the overheat protection can be switched off in the service menu;
overheating can cause damage to the optical system.