      Ref. out of range Reference out of range.
      The counter of the internal
      reference capacitor (signal
      conversion) is not within
      the expected limits.
      Parameters wrong.
      Switch off the instrument,
      wait one minute then
      switch the instrument on.
      Check parameter settings.
      Call service technician.
      Check that the offset setting in the temperature menu is the same as in
      the instructions on the photometer board.
      Refer to chapter 8.5.2, "TEMPERATURE category settings" to carry out
      a functional check of the temperature controller and sensor offset set-
      tings.
      Replace the photometer board, if necessary.

  • @محمدالحمادي-ز5ف
    @محمدالحمادي-ز5ف 2 місяці тому

    لوسمحت اخي ماهو سبب هذي المشكلة ERRORCODE:0 لو سمحت أكتبلي الحل في التعليق

    • @tech.keys.
      @tech.keys.  2 місяці тому

      Flow cell did not reach
      37 °C +/- 0.2 °C (offset is
      taken into account) within
      1 hour after start up.
      Peltier element/sensor
      defect.
      Note: a test is made before
      every measurement to
      check that the temperature
      is stable at 37 °C +/- 2°C.
      If the temperature is out of
      range, a T-flag is added to
      the results.
      Note: when a temperature
      error occurs the
      peltier element is switched
      off to prevent damage to
      the flow cell.

    • @tech.keys.
      @tech.keys.  2 місяці тому

      Switch off the instrument,
      wait one minute then
      switch the instrument on.
      Check that the environ-
      ment conditions are within
      the specified limits.
      Check that the filter at the
      rear of the instrument is
      not clogged.
      Ensure that the environ-
      ment conditions are met.
      Call service technician

    • @tech.keys.
      @tech.keys.  2 місяці тому

      Check temp sensor. If the sensor is malfunctioning replace the photo-
      meter board; remember to set the offset in the temperature menu.
      Check resistance of the peltier element. Replace if necessary.
      Check connections between the processor board and the peltier ele-
      ment/sensor (please refer to the Appendix, A.4 "Electrical Drawings").
      Check connections to the integrated photometer board are secure.
      Check for obstructions in the airflow to the heat sink of the peltier
      element.
      Check temperature paste between heat sink peltier and house assy.
      Check that the offset setting in the temperature menu is the same as in
      the instructions on the photometer board.
      Refer to chapter 8.5.2, "TEMPERATURE category settings" to carry out
      a functional check of the temperature controller and sensor offset set-
      tings.
      Note: the overheat protection can be switched off in the service menu;
      overheating can cause damage to the optical system.

    • @محمدالحمادي-ز5ف
      @محمدالحمادي-ز5ف 2 місяці тому +1

      @@tech.keys. شكراً لك من قلبي أخي على هذا المجهود الذي فعلته معي وايضاً على تواضعك واخلاقك الرائعة في معاملتك مع من يطلب العلم حقاً انت انسان رائع وجيد

    • @محمدالحمادي-ز5ف
      @محمدالحمادي-ز5ف 2 місяці тому +1

      @@tech.keys. شكراً لك أخي على مجهودك حقاً انت إنسان رائع

  • @mukeshkumarmadhur9224
    @mukeshkumarmadhur9224 9 місяців тому

    Hi

  • @WaqasAhmad-yf8fx
    @WaqasAhmad-yf8fx 5 місяців тому

    Sir apna contact number share kar dian ap sa board rapair karwana ha Microlab 300 ka same issu ha tamparture ka