EDS analysis on Tescan SEM

Поділитися
Вставка
  • Опубліковано 1 гру 2024

КОМЕНТАРІ •

  • @felixugartecejas1878
    @felixugartecejas1878 2 роки тому

    Excellent series of videos, really useful.

  • @zellaparry
    @zellaparry 3 роки тому +1

    Excellent series of videos, very useful.
    If you were to coat a sample with gold, how can you eliminate the gold data from your EDX results?

    • @michigancenterformaterials5020
      @michigancenterformaterials5020  3 роки тому

      Hi Zella, there are software tools available such as DTSA-II to remove the gold peaks if you have gold coated your sample, but we find it best to leave the peaks in the spectrum and simply label them. Gold coating is an accepted part of electron microscopy and if any sort of x-ray quantification is done absorption by the Au will need to be accounted for.

  • @felixugartecejas1878
    @felixugartecejas1878 2 роки тому

    The third column appears , it is standard deviation or percentage error?

  • @mohsenkhalilinajafabadi7195
    @mohsenkhalilinajafabadi7195 4 роки тому

    It was so useful. Thanks a lot

  • @AnangKatyayanrandomvideos
    @AnangKatyayanrandomvideos 4 роки тому

    very useful thanks a lot

  • @gamal33h
    @gamal33h 11 місяців тому

    Please i need ti contact with you regarding to a problem in our EDAX Team