Change the Split Vent Trap - GC Troubleshooting Series

Поділитися
Вставка

КОМЕНТАРІ • 7

  • @borkbirk3272
    @borkbirk3272 7 років тому +3

    Great video!

  • @mohamedmamdouh4782
    @mohamedmamdouh4782 4 роки тому

    Hi
    Could you plz help me relating to GC valve box temperature that not return to set temperature and I had to make air connection to cool it up to reach set point.

  • @charlesp6461
    @charlesp6461 2 роки тому

    Is there any need to do this if you only run in pulsed splitless mode?

  • @carlag07
    @carlag07 5 років тому

    Hello, May i know how much volume that i need to inject because my sample is on gas manual injection ?Do u have any method or any reference or video that i can refer for...Thank you

    • @agilent
      @agilent  5 років тому +1

      The volume of gas/sample you inject will vary based on your instrument parameters, analyte concentration, and the type of detector you are using. 1 mL or less of headspace may be fine for some applications, but other times 2-5 mL injections may be needed. This has a lot to do with what compounds you are looking at, and how easily they move into the gas phase. Additional videos you may want to watch: 7697 Headspace Sampler www.agilent.com/en/products/gas-chromatography/sample-preparation-introduction/7697a-headspace-sampler/7697aheadspacesampler-video and 7697A Headspace Sampler Method Development Wizard www.agilent.com/en/video/7697a-headspace-sampler-method-development-wizards-video or call Agilent Tech support www.agilent.com/en/contact-us/page for guidance.

  • @arunkumar-bj2hi
    @arunkumar-bj2hi 6 років тому

    when we have to replace

    • @agilent
      @agilent  6 років тому

      This depends on a few factors. In general it is suggested to replace the Split Vent Trap at least once a year. But if you are running Split injections with high split ratios, and are seeing ghost peaks in blank injections, then you may want to change it more frequently.