IMRT Concept

Поділитися
Вставка
  • Опубліковано 18 гру 2024

КОМЕНТАРІ • 6

  • @sumeerofficial5692
    @sumeerofficial5692 Рік тому +1

    Thanks for the upload.. may I know what FS you went down for small field.. did you treat SRS case 1cc lesion volume.?

    • @MPTFoundation
      @MPTFoundation  Рік тому

      Well smallest we treated was 0.7 cc on true beam with 0.5cm MLC width

    • @sumeerofficial5692
      @sumeerofficial5692 Рік тому

      @@MPTFoundation did you measured output factor for 5 mm FS ?

    • @MPTFoundation
      @MPTFoundation  Рік тому

      @@sumeerofficial5692 0.6 mm least verify with TPS calculated value

    • @sumeerofficial5692
      @sumeerofficial5692 Рік тому

      @@MPTFoundation thanks for the reply.. may I know your TPS? As far as I know eclipse will only accept 1x1 cm FS data for beam model configuration for output factor table.. so below 1 cm target how will you know the TPS calculated one is true value or not.. expect we can verify TPS with some end to end test.. did you do any special test addressing this matter?

    • @MPTFoundation
      @MPTFoundation  Рік тому

      @@sumeerofficial5692 Watch this ua-cam.com/video/cgoezwfYGc4/v-deo.html i have explained in this video