BME RHC Training of SERVO i Ventilator (Maquet)

Поділитися
Вставка
  • Опубліковано 12 вер 2024
  • Overview of ventilator.
    Basic Connections of ventilator.
    Preuse check
    Expiratory Cassette

КОМЕНТАРІ • 18

  • @ajitprajapati1823
    @ajitprajapati1823 4 місяці тому

    Very nice information

  • @arpita4984
    @arpita4984 3 роки тому +3

    Very nice sir

  • @flower14356
    @flower14356 2 роки тому +1

    Clearly understood thank you so much sir your uploadeding this video

  • @kyleweis2724
    @kyleweis2724 3 роки тому +1

    Good video but you bypassed the patient circuit test! It said to connect the patient circuit and you just used the single tube. If there is a leak in the circuit you would catch that with the patient circuit test. This is very important to do correctly!

  • @Parker_Prasad
    @Parker_Prasad Рік тому

    Very informative sir

  • @srutia425
    @srutia425 4 роки тому

    Very nice

  • @adarshsrivatsasa706
    @adarshsrivatsasa706 3 роки тому

    Please doo more videos

  • @rekhuthakur6006
    @rekhuthakur6006 2 роки тому

    Pls help we try callibration but shows internal leakage or exp. Cassette

    • @nandkumarjambhale2287
      @nandkumarjambhale2287 2 роки тому

      There are number of possibility of internal leakage test....
      1) change expiratory cassette and confirm ....the problem with cassette or ventilator
      2) if problem with cassette then open membrane and clean it and fix it properly. Your problem will resolve.
      3) if problem with ventilator then you have to check one by one parts

    • @bmerubyhallclinic8519
      @bmerubyhallclinic8519  2 роки тому

      your problem resolved??? you can call me 7588342291

  • @flower14356
    @flower14356 2 роки тому

    Belevestor ventilator settings video apload please

  • @girishshidore8029
    @girishshidore8029 4 роки тому

    Very nice sir