Medical Gas Brazing Med-gas outlets in MRI -How to braze properly with no leaks

Поділитися
Вставка
  • Опубліковано 20 жов 2024
  • We were asked to do the inwall and get inspection before connecting to the main.
    We don’t normally do this but this what we came up with as we had to purge backwards!

КОМЕНТАРІ • 15

  • @jonathanzepeda2482
    @jonathanzepeda2482 Рік тому +5

    When working with MedGas outlets it’s always best to wrap them with wet rags in order to avoid heat damage to the o rings it has inside. Prevents leaks in the future when you pressurize or install finish.

  • @Martinhockey12
    @Martinhockey12 3 місяці тому

    love the content brother, should post more often!! what state is this?

  • @GP24777
    @GP24777 3 місяці тому

    Where do you get that adapter from to go into the outlet?

  • @robmay2489
    @robmay2489 6 місяців тому

    You should always build everything out and test it first. Tie-in to the main should be the last thing you do without a purge, never purge against a system! I'm assuming this is a remodel/addition. If it's all new construction please feel free to tell me to zip it!

    • @noahbell7297
      @noahbell7297 3 місяці тому +1

      They’re installing Oxygen, Medical air, Medvac, and Waste anesthetic gas disposal systems. They’re using nitrogen to displace all of the oxygen out of the system so they can braze the correct way. After this they’ll do their initial blowdown and then their purge test. They’re not done yet lol. This is the textbook way to braze med gas pressurized/vacuum systems.

    • @zackhurley9148
      @zackhurley9148 2 місяці тому

      ​@noahbell7297 is there a book to learn more on step by step when installing zones,and outlets and purging steps that could be looked up ?

    • @robmay2489
      @robmay2489 2 місяці тому

      ​@@noahbell7297 I thought I'd read somewhere that the initial blow down gets done before the installation of rough inlet/outlet bodies? I'm getting old, my memory must be slipping! My personal preference would be to not blow any debris through those o-rings and poppets.

    • @noahbell7297
      @noahbell7297 2 місяці тому

      @@zackhurley9148 NFPA 99 medical gas guidelines, I have the books at home I’ll reply back later for the exact title.

    • @noahbell7297
      @noahbell7297 2 місяці тому

      @@robmay2489 you’d be correct, personally i do both. Before and after installation just in case. Even if you do a repair on a singular fitting it needs to be blown / purged again

  • @josephrodriguez4542
    @josephrodriguez4542 2 роки тому

    how many pounds of flow do u keep flowing through while brazing?

    • @rodgraff1782
      @rodgraff1782 Рік тому

      No pounds, just flow

    • @medgaslife4509
      @medgaslife4509  Рік тому +1

      @@rodgraff1782 depends on how many open ends or outlets,usually 5-10 cfm on the flow meter

  • @osslv
    @osslv 2 роки тому

    Is that torch's tip custom made?