Liquid Penetrant Testing

Поділитися
Вставка

КОМЕНТАРІ • 98

  • @dan-xl4mg
    @dan-xl4mg 3 роки тому +18

    this is so much easier to understand then my teachers explains it in front of 30 students, so thank you for this amazing video, I will definitely recommend this to the other students.

  • @abc6312
    @abc6312 Рік тому +13

    If provide more close-up pictures for all indications on the second test plate, it will be great.

  • @nomuplays7487
    @nomuplays7487 5 місяців тому

    Why u did to and fro motion which doing solvent removal? It's clearly written in ASME BPVC SEC 5 ARTICLE 6

  • @sv7382
    @sv7382 9 років тому +1

    Great demonstration Robert. Mike told me the story behind the test plate. Sneaky!

  • @irfanakram3044
    @irfanakram3044 28 днів тому

    Nice explanation

  • @nathanbra8605
    @nathanbra8605 5 місяців тому

    If you spray a paint brush with penetrate and brush weld, will that decrease results or is that ok?

  • @The3Shogun
    @The3Shogun 9 років тому +16

    You must wipe off pentrant vertically or horizontally only. One direction. And way to much developer on your HAZ.

    • @robertmertz7544
      @robertmertz7544 8 років тому +3

      +The3Shogun You are correct for most PT applications regarding cleaning. I disagree about the HAZ. I suppose I could have masked the area of base metal that was not under inspection but the developer spray covers a large area and cleans off very easy. The procedure I was following allows me to wipe in both directions and as indicated I won't have a lot of non-relevant indications that need to be re-tested. Also, the weld and HAZ are evaluated based on VT after PT so I can wipe penetrant out of U/C, etc. and measure for acceptance. Cracks that are very tight would not be wiped away but may be missed by VT. I could have explained why I was cleaning in both directions. Thanks for the feedback.

  • @kalpeshsingh2670
    @kalpeshsingh2670 3 роки тому

    How to find out the length of a circular defect whose end is loose? as in not connected?

  • @FacetiousUniqueness
    @FacetiousUniqueness 8 років тому

    Excellent Video! Thanks!

  • @riddhisarvaiya5243
    @riddhisarvaiya5243 6 років тому

    You should clean the plate in one direction. And also apply penetrant and developer in one direction. It is either horizontally or vertically. Not both direction you can use for perfect results.

    • @newmovieschannel1364
      @newmovieschannel1364 6 років тому

      ya mam absoultly write ...he is not correct because of pentrates also pick out in defects

    • @newmovieschannel1364
      @newmovieschannel1364 6 років тому

      ya mam absoultly write ...he is not correct because of pentrates also pick out in defects

    • @razvancampean
      @razvancampean 5 років тому

      can you specify where does it say so? in which standard or norm (what page)?

  • @DPeter1967
    @DPeter1967 10 років тому

    Great Jog Robert! Haven't seen you since the New Bay Bridge Project.

  • @georgevitor6975
    @georgevitor6975 3 роки тому

    After mechanical cleaning ( example wire brushing,grinding,buffing) of part to be tested,is cleaning of part by means of cleaner/remover still necessary?or it can be skipped and apply penetrant already?

  • @ryanarnold2813
    @ryanarnold2813 7 років тому

    Very Nice, Well Done !

  • @حسنعليعبيدعوض
    @حسنعليعبيدعوض 3 роки тому

    معهد النجف 😎 يرحب بكم

  • @shreeramvenkatesan1658
    @shreeramvenkatesan1658 4 роки тому

    This guy is really really cool

  • @alleneasomathew
    @alleneasomathew 2 роки тому

    Keep doing

  • @givenmahlangu1266
    @givenmahlangu1266 5 років тому

    Is Post cleaning necessary ?

    • @rubenrosales7823
      @rubenrosales7823 4 роки тому

      Given Mahlangu depends where you are and who you are working for. My preference is to leave clean, unless there are any relevant indications. If there is a defect that is Rejectable I mark up and leave so the welder/QC or whoever is the one you are reporting to may see... I have worked places where they are big on us leaving it clean and also have been places where they just go to next job/weld...

  • @samueljacob6031
    @samueljacob6031 5 років тому

    thank you sir

  • @itsimzaa621
    @itsimzaa621 Рік тому

    Will the fluorescent non-aqueous developer produce better results in evaluating the defects than the visible non-aqueous developer?

  • @hectorpaniagua5702
    @hectorpaniagua5702 6 років тому

    What method is this

    • @spoonbill01
      @spoonbill01 6 років тому

      Hector Paniagua This is Type 2 (visible ) Method c ( solvent removeable ) Form e ( nonaqueous type 2 visible dye solvent based ) Class 2 solvent ( non-halogenated )

  • @adamismail9895
    @adamismail9895 8 років тому

    super

  • @chanthasavann
    @chanthasavann 4 роки тому

    Hello Guy let mark your Friend too

  • @bn70066
    @bn70066 9 років тому

    hi am NDE man from middle east :))

  • @ar-1571
    @ar-1571 Рік тому +1

    Great video! My teacher played this during class and me and my friends laughed at this because you made it look like a Porn video.. I mean the way you were breathing and cleaning 🤣🤣

  • @victornjoku9684
    @victornjoku9684 2 роки тому +1

    Why cleaning zig-zag.,? Clean from one side.

  • @marcodiegocambronerovillal7647
    @marcodiegocambronerovillal7647 3 роки тому +9

    Me studying, getting asmr vibes from this lmao

  • @kmondak3387
    @kmondak3387 8 років тому +8

    Should have tilted the plate to do away with glare and let the viewer see. An intentional crack in the plate would have been good for checking engine heads etc.

  • @dginia
    @dginia 3 дні тому

    How about inside the cylinder of an aluminum engine block with just the head removed?

  • @srinivasandwarakanathan3822
    @srinivasandwarakanathan3822 2 роки тому +1

    Well explained, But as per standard practice while applying , cleaning pentetrant from the surface we should not wipe or clean bi-directionally. It shall be cleaned only in one direction, the same is applicable while applying penetrant. Hope i am correct.
    Regards
    Dwarakanathan ASNT LEVEL III

  • @planetbeatmobb1762
    @planetbeatmobb1762 8 років тому +8

    great job, we were taught to spray the developer away from the metal first to make sure it's fully agitated tho?

    • @ifaheemonline
      @ifaheemonline 8 років тому

      +Planet Beat Mobb Yes you are right. it ensures the uniform spray of developer.

  • @writer1meb
    @writer1meb 7 років тому +4

    Thank you. This video is very helpful. We didn`t know before watching this video how to use liquids.

  • @jjjsss3869
    @jjjsss3869 8 років тому +27

    A close up of the plate with the defects so we could actually see what they look like would have been helpful.

  • @arvindnavinsv2227
    @arvindnavinsv2227 3 роки тому +3

    thanks man you did a great job explaining, i am gonna show this to my student for easy understanding

  • @shanwarn3772
    @shanwarn3772 6 років тому +7

    Hi sir
    good morning
    I'm from india
    i finished NDT Level 2
    i need a job
    did any vacancy is in your factor

  • @socialhelp5704
    @socialhelp5704 6 років тому +1

    removing pen
    etrant is only in one direction ... To and fro is not allowed

  • @dishitapurohit2478
    @dishitapurohit2478 9 років тому +3

    Can you provide with the product name and specifications of the red spray dye and cleaning sprays you used? also can this be used to test for solders on PC boards

    • @robertmertz7544
      @robertmertz7544 8 років тому

      +Dishita PUROHIT I would recommend using RT for PC boards.

  • @akkusrocks3367
    @akkusrocks3367 7 місяців тому

    Sir penetrant cleaning need to do in one direction , am i right ?

  • @brettb.7425
    @brettb.7425 5 років тому +1

    Good demonstration but please get the camera closer to show the details. Thank you for posting this!

  • @tradingmachine65
    @tradingmachine65 3 роки тому

    make sure cleaner flashes off for correct time, 5 minutes min. old cleaners used to flash much faster. if cleaner remains in crack, inspection is void, if being audited you'll be flagged

  • @3sempravin.r13
    @3sempravin.r13 3 місяці тому

    It's cleaning process is not accepted

  • @brianb5779
    @brianb5779 3 роки тому

    Camera guy afraid to get close. Otherwise good info

  • @esmeraldaalviso2220
    @esmeraldaalviso2220 3 роки тому

    I didn't like it cause a indication in the plate will be better explanation.

  • @finessedagain4557
    @finessedagain4557 4 роки тому +1

    You are an absolutely legend, I hope you know that pal

  • @fahdhanif4150
    @fahdhanif4150 3 роки тому

    Watching this just for exam purposes 😑😑😑

  • @rmvmedia4897
    @rmvmedia4897 5 років тому +1

    Wrong apply method

  • @yuvrajbhatt1050
    @yuvrajbhatt1050 6 років тому +3

    It was nice illustration of the whole process. Can u tell which cleaner, dye & developer you used here?

    • @prasadsawant1213
      @prasadsawant1213 3 роки тому +1

      Pmet is one the oldest and leading companies.
      Different products available as per different sizes of cracks

  • @ДианаНиколаева-ы3т

    CovidImages need to be invested more than half19

  • @crackersphdinwumbology2831
    @crackersphdinwumbology2831 6 років тому +1

    I'm an Air Force NDI technician.
    It's kind of shocking how similar yet different Method C inspections are in the civilian sector. Only 20 minutes of dwell time only using visible dye penetrant! It makes sense though, given its easier, faster, and it's doubtful you need as much sensitivity.

    • @subinperingave9777
      @subinperingave9777 6 років тому

      Hi..
      I have a doubt.
      Is there any need to do penetrate test on a complete machining part ?

    • @chuckb9595
      @chuckb9595 2 роки тому

      That was only 10 min of dwell time.

  • @wizyasilupumbwe2497
    @wizyasilupumbwe2497 7 років тому +1

    thanks am actually writing an end of term essay on this, cleared alot of things up thanks.

  • @ptrangdang
    @ptrangdang 7 років тому +2

    very clear. well-done. thank you

  • @mood666
    @mood666 Рік тому

    6:00

  • @hahaclintondix6099
    @hahaclintondix6099 2 роки тому

    This is definitely Walter white Heisenberg

  • @muhammadahsanmehar3265
    @muhammadahsanmehar3265 4 роки тому

    Why we are not use to cleaner direct sprey in welding .

  • @TheAjnett23
    @TheAjnett23 5 років тому

    your camera man/women is not so very good..

  • @tanveerahmed3094
    @tanveerahmed3094 8 років тому +1

    thanks very helpful.

  • @ezekielmunatsi5496
    @ezekielmunatsi5496 2 роки тому

    Very informative, Thank you.

  • @abidsaleemdesconian3528
    @abidsaleemdesconian3528 4 роки тому

    Thanks for fine explaination.

  • @jithinmathewjoseph9465
    @jithinmathewjoseph9465 4 роки тому

    can this method be used in pipes

  • @raymonddupuis9461
    @raymonddupuis9461 9 років тому

    Nice video moderately informal though....

  • @rubenrosales7823
    @rubenrosales7823 4 роки тому

    Close up of defects?

  • @adilhafeez5942
    @adilhafeez5942 5 років тому

    Really good explanation thx

  • @imanebahar286
    @imanebahar286 4 роки тому

    great video

  • @malikahmad781
    @malikahmad781 5 років тому

    superb video

  • @narendernayakguguloth5018
    @narendernayakguguloth5018 6 років тому

    Tq sir

  • @yacinebenz1349
    @yacinebenz1349 9 років тому

    great job

  • @Gabby-du4mc
    @Gabby-du4mc 6 років тому

    Thank you!!

  • @abdelwahedbanouji2638
    @abdelwahedbanouji2638 8 років тому

    i wanna learn the liquid penetrant teasting cuz im a radio log i wanna some info about that and i wish if someone contac me for ask him meny thing about liquid and magnet

  • @redcarpet2544
    @redcarpet2544 2 роки тому

    Hello! Thanks for useful video! I have the only one question...Why don't you use a respirator? These sprays are extremely harmful for your health, especially for lungs.

  • @kuranenuhaa1880
    @kuranenuhaa1880 2 роки тому

    Well Explained. Thank you sir, please can u list some examples of the liquid penetrant commonly used

  • @saqarjamel3383
    @saqarjamel3383 6 років тому

    Goog

  • @somnathghosh9873
    @somnathghosh9873 9 років тому

    Post cleaning is not proper Way..

    • @JavierD
      @JavierD 9 років тому +1

      Somnath Ghosh why?

    • @somnathghosh9873
      @somnathghosh9873 9 років тому

      for post cleaning you must clean the surface, from first point to 2nd point you can't clean first to 2nd and again 2nd to 1st. you can clean any of one way 1st to 2nd or 2nd to 1st jv071

    • @chevellecarcass
      @chevellecarcass 9 років тому

      Somnath Ghosh Post cleaning after the penetrant is applied? What do you mean by 1st to 2nd?

    • @JavierD
      @JavierD 9 років тому

      ***** 1st point to 2nd point. What Somnath state is that one should clean in the same direction, always right to left (for example)

    • @somnathghosh9873
      @somnathghosh9873 9 років тому

      yaa you are right