  • Oscilloscope probes are an essential extra to any oscilloscope, but to get the best performance from the probe and hence the scope, the probe compensation must be correctly adjusted.
    Errors in the the scope probe compensation adjustment will result in very noticeable errors when making measurements, so it is absolutely necessary to get the compensation adjustment right.
    This video describes which oscilloscope probes need compensation adjustment and why this is needed. The video also shows how to adjust the scope probe compensation and details of the correct tools.
    Understanding why scope probe compensation is needed, helps see the need for accurately compensating probes each time they are used. It can also help explain the reasons for gaining poor results on the occasions when anyone forgets to compensate the probe.
    There are many different varieties of oscilloscope probe, but the type most commonly used is the passive voltage probe. Of these the x1 or times one probe provides a straight through connection and this enables the input impedance of 1 Mohm to be seen at the probe tip. (There are some 50 ohm probes but these are used for specialist RF applications and are not particularly common).
    The times ten probe or x10 probe has a series 9 Mohm resistor and gives an overall input impedance of 10 Mohms, although this is at DC and low frequencies. As the frequency rises the scope input capacitance increasingly affects the input impedance.
    To give a flat response a compensation capacitor is added in the probe. This is adjusted to give the correct potential divider effect. If the probe is under compensated then square waves will appear rounded and if over compensated a square wave will have overshoot on the leading edge.
    Getting the oscilloscope probe compensation right will enable waveforms to be seen accurately and accurate measurements to be made.
    More information about oscilloscope probes and scope probe compensation: www.electronics-notes.com/art...
    Check out the Electronics Notes website: www.electronics-notes.com
    Please don't forget to subscribe to our UA-cam channel: ua-cam.com/users/electronicsnotes?...
  • Наука та технологія

КОМЕНТАРІ • 16

  • @markharrisllb
    @markharrisllb 2 роки тому +1

    Wow, that was clear and easy to understand, thank you very much!

  • @SvenSchumacher
    @SvenSchumacher 2 роки тому

    Simple to do, but really important! Thanks!

  • @uiyrf
    @uiyrf 2 роки тому

    very thoroughly and helpful instruction

    • @ElectronicsNotes
      @ElectronicsNotes  2 роки тому +1

      Glad you liked the video & found it useful. We try!

  • @ottoleroy2001
    @ottoleroy2001 7 місяців тому

    very nice video!

  • @TheNoCodeTech
    @TheNoCodeTech 2 роки тому +1

    Very helpful thanks

    • @ElectronicsNotes
      @ElectronicsNotes  2 роки тому +2

      Really glad it helped you. helps make the time we spent preparing it more worth while.

  • @ytubeleo
    @ytubeleo 2 роки тому

    Great video

    • @ElectronicsNotes
      @ElectronicsNotes  2 роки тому +2

      Thank you very much for your comment. Greatly appreciate it.

  • @bluetorch13
    @bluetorch13 Рік тому

    I though I damaged mine for using a metal screwdriver. This video saved me! lol.

  • @ramiroruizdecastilla3085
    @ramiroruizdecastilla3085 9 місяців тому

    Hello. What about automotive oscilloscope probes? They have no slot for adjustment. That means they cannot be compensated? They come with 2 alligator clip ends or banana plugs but no slot anywhere.

    • @ElectronicsNotes
      @ElectronicsNotes  8 місяців тому

      It is probably because the probes are x1 only, although I am not familiar with automotive probes. It is only where you have x10 or greater that the compensation is required.