How to Change an HPLC Rotor | Shimadzu SIL-20A

Поділитися
Вставка
  • Опубліковано 7 лют 2025

КОМЕНТАРІ • 11

  • @samanthac323
    @samanthac323 2 роки тому

    Awesome video and very helpful. Thank god I'm left handed. 👍

  • @gharibabdoo3945
    @gharibabdoo3945 7 місяців тому

    Thank you 🎉 David , always great

  • @shahidullah233
    @shahidullah233 8 місяців тому

    your tutorials are amazing. would make a video for LPV rotor & stator replacement ??

  • @geekyogurt1904
    @geekyogurt1904 11 місяців тому

    I think it would be a good idea to say when you might want to consider changing the rotor. i.e. what kind of problems would you be having that would make you think the rotor needed changing

    • @DavidYazdi1
      @DavidYazdi1  11 місяців тому +1

      This is a very good comment. Reasons would be leaking or inaccurate injection volume dispensing.

  • @shuhaidashukri2299
    @shuhaidashukri2299 Рік тому

    hi, is it need to do hpv rotation or hpv check after the inspection or reassemble? since im faced with high preesure (max sett 15mpa) when rinsing or zhome. plus no drain out during rinsing or autosampler purge.

    • @DavidYazdi1
      @DavidYazdi1  11 місяців тому

      2000 psi isn't really that high. But if you have no column, it's super high. You should be checking out the valve with just a union installed so you can get a good baseline for pressure without the column.

  • @shuhaidashukri2299
    @shuhaidashukri2299 Рік тому

    do we need to do hpv rotation or hpv check after the reassembly? im facing high pressure when needle rinse or Z home . Plus there is no fluid drains out when im purging the autosampler or rinse needle.. max preesure set is 15mpa.

    • @IvanGonzalez-bn1yj
      @IvanGonzalez-bn1yj Рік тому

      were you able to solve this? I am having the same issue

    • @DavidYazdi1
      @DavidYazdi1  11 місяців тому

      Did you check to see if there is a clogged tubing? And if you're purging the autosampler, the liquid will come out of the drain tubing.