  • When performing a split injection, there are several considerations to keep in mind to improve the quality of your chromatography. But what should you change and why?
    In this Restek Tip, we’re going to show you how to optimize your split injections. We’ll start with liner selection, before discussing the importance of split ratios. Next, we’ll focus on optimizing the inlet temperature, and finally, the initial oven temperature.
    -----------------------------------------------------------------
    Additional Resources
    GC Inlet Liner Selection: Split Liners, Part 1 | www.restek.com/chromablograph...
    GC Inlet Liner Selection: Split Liners, Part 2 | www.restek.com/chromablograph...
    How to Choose a GC Inlet Liner | www.restek.com/articles/how-t...
    GC Inlet Maintenance Guide | www.restek.com/articles/gc-in...
    -----------------------------------------------------------------
    Restek Recommended Tools & Parts
    Restek Inlet Liners | www.restek.com/c/1210
    Restek Electronic Leak Detector | www.restek.com/p/28500
    -----------------------------------------------------------------
    Connect with Us
    Comment on this video and subscribe to Restek's UA-cam channel today at / restekcorp
    Website: www.restek.com
    LinkedIn: / restek-corporation
    Facebook: / restekcorporation
  • Наука та технологія

КОМЕНТАРІ • 10

  • @michaelbalagula7419
    @michaelbalagula7419 Рік тому +1

    High quality series, both in explanation and animation! Thank you a lot

  • @justforentertainment74
    @justforentertainment74 2 роки тому +1

    Thank you for explaining

  • @hanumantharaorao5874
    @hanumantharaorao5874 Рік тому +1

    Thank you sir

  • @meme2403
    @meme2403 7 місяців тому

    You talk about the initial oven temperature, does this mean that the separation was carried out in temperature programming mode? Tank you.

    • @RestekCorp
      @RestekCorp  7 місяців тому

      Generally yes, but it doesn't need to be.

  • @quimicamentefalando
    @quimicamentefalando Рік тому

    Congratulations for the video, but I don't understand how a pulsed split injection can have the liner with increased pressure if the injector outlet opens?

    • @RestekCorp
      @RestekCorp  Рік тому +2

      The pressure controls of the GC system are designed to handle these pressure changes. So while there is still flow going through the column (and out the split vent, septum purge, etc.) pulsed injection allows users to temporarily elevate the pressure in the injection port.

    • @quimicamentefalando
      @quimicamentefalando Рік тому +1

      @@RestekCorp Great! Thank you for this.

    • @RestekCorp
      @RestekCorp  Рік тому

      You're welcome!