USP General Chapter 41 || Balances || Part-2 ||

Поділитися
Вставка
  • Опубліковано 6 вер 2024

КОМЕНТАРІ • 48

  • @gooselukas1317
    @gooselukas1317 Рік тому +2

    This was just what I needed to fill the gap in my knowledge. Thanks a lot!

  • @vijayparmar5455
    @vijayparmar5455 2 роки тому +1

    Superb so good like this video 💯👍🙏🙏

  • @Sambhjinagar
    @Sambhjinagar 2 місяці тому

    1:19 good

  • @ramakrishnaguttula7941
    @ramakrishnaguttula7941 2 роки тому +2

    Good information sir 👍

  • @rajuahmmed5882
    @rajuahmmed5882 Рік тому +1

    ❤❤❤❤

  • @lekang8179
    @lekang8179 Рік тому +1

    thank you for summarizing this information!!

    • @PharmaPill
      @PharmaPill  Рік тому

      Glad it was helpful!

    • @rabiierraji3036
      @rabiierraji3036 8 місяців тому

      ......
      😅​@@PharmaPill😮😮😮😮😮😮

  • @sanjays1344
    @sanjays1344 Рік тому +1

    Thank you informative

  • @shahid_palguni1
    @shahid_palguni1 2 роки тому +1

    Thank you for this video

  • @sydneysantiago974
    @sydneysantiago974 2 місяці тому

    Thank you!!

  • @manojmodi414
    @manojmodi414 2 роки тому +1

    Thank you very much sir

  • @solution24x50
    @solution24x50 Рік тому +1

    Hi Sir Please make video on Dissolution performance verification using USP Prednisone tablet..I have gone through certificate but not able to understand

  • @surendrababuchallagundla9249
    @surendrababuchallagundla9249 2 роки тому +1

    Please do one video on reviewed oos guideline

  • @santhoshvarma9339
    @santhoshvarma9339 Рік тому +1

    Sir as per the formula if I perform, the answer is always 0.10
    For suppose
    (2xsdx100)/2000xsd=0.10
    Am right or wrong

    • @gayathribabu1214
      @gayathribabu1214 19 днів тому

      Yes exactly...always it's 0.1% then what is the right formula

  • @ajitjadhav607
    @ajitjadhav607 2 роки тому +2

    Is d value & least count is same ?

  • @vijaybirhade9424
    @vijaybirhade9424 Рік тому +1

    Why we are selecting 5 gm weight for repeatability test on analytical Balance? Plz tell me about this sir.🙏

    • @PharmaPill
      @PharmaPill  Рік тому

      Just check is it 5% of your balance capacity? If answer is yes then you have selected this weight based on the European Pharmacopeia. As per EP to carry out Repeatability test, weight should be in lower end and should not more than 5% of balance capacity.

  • @muralibond1373
    @muralibond1373 Рік тому +1

    For rough top pan balance for monthly verification is only accuracy enough or we need to perform repeatability also..

    • @PharmaPill
      @PharmaPill  Рік тому +1

      As per European pharmacopeia Repeatability and Sensitivity.
      As per USP All 5 parameters required.

  • @ajitjadhav607
    @ajitjadhav607 Рік тому +1

    Sir, For accuracy test only weight shall be use for calibration or all weights shall be use for calibration?

    • @PharmaPill
      @PharmaPill  Рік тому

      Only one test weight is taken for accuracy test. Generally between 5 to 100 % of balance capacity weight is taken.

    • @ajitjadhav607
      @ajitjadhav607 Рік тому

      Sir as per accuracy formula, if observed weight will get higher value than certified value then calculation results get in minus value, how we can write final results (limit is NMT 0.10 %)

    • @ajitjadhav607
      @ajitjadhav607 9 місяців тому

      Hello sir,
      Hope you are doing well.
      Let me know that a statement given in accuracy parameter that ' the test weight's max. Permissible error (mpe), or alternatively it's calibration uncertainty shall be NMT one third of applied test limit of the accuracy limit. Plz explain .

  • @ajitjadhav607
    @ajitjadhav607 2 роки тому +2

    Please explaine, In repeatability there is one note that is 'The test weight must be within the balances capacity, but the weight need not be calibrated.'

    • @PharmaPill
      @PharmaPill  2 роки тому

      Test weight must be within in operating range not balance capacity…

    • @ajitjadhav607
      @ajitjadhav607 9 місяців тому

      Hello sir, Hope you are doing well.
      Sir let me explain below;
      Repeatability formula = 2000 x SD / Smallest net weight
      Repeatability= 2000 x SD x 100 / smallest net weight
      Which one is correct and how ?

  • @dashsubhasish5667
    @dashsubhasish5667 4 місяці тому

    What is certificate value

  • @rameshmanda3877
    @rameshmanda3877 Рік тому

    Sir. Why 2000 we are using .. why not use for 1000 could you please explain to us

  • @ajitjadhav607
    @ajitjadhav607 2 роки тому +1

    Sir, if i perform repeatability parameter as per monthly calibration then there may be possibility of changes in SD every time. In this case how we can find out smallest net weight of balance ?

    • @PharmaPill
      @PharmaPill  2 роки тому

      Smallest net weight need to change based on your monthly Repeatability test calibration. But ideally you will see there will not be much change in smallest net weight because most of the time your SD will be less that 0.41 d

    • @ajitjadhav607
      @ajitjadhav607 2 роки тому +4

      Ok sir.
      i have a 5 digit semi micro balance with range of 1 mg to 200 gm. As per USP chapter we can calculate smallest net weight by using formula 2000 x 0.41d i.e. 2000 x 0.41 x 0.00001 = 8.2 mg.
      Then can i weight below 8.2 mg for any HPLC analysis.
      Most of the times we have very less quantity of CRS or Impurity standard so In this case, can we weight below 8.2 mg for HPLC or other analysis preparation?

    • @PharmaPill
      @PharmaPill  2 роки тому +1

      @@ajitjadhav607 excellent Question. Don’t use for weighing below 8.2 mg. You should use ultra micro balance or micro balance.

    • @bestisyettocome2645
      @bestisyettocome2645 Рік тому

      @@PharmaPill sir it means ye 8.2 least count hai
      How to calculate least count of weighing balance. Plz explain

    • @bestisyettocome2645
      @bestisyettocome2645 Рік тому

      @@PharmaPill sir it means ye least count h ??

  • @bhuvip9204
    @bhuvip9204 Рік тому +1

    How to find minimum load of balance?

    • @PharmaPill
      @PharmaPill  Рік тому

      2000xSD is the formula

    • @bhuvip9204
      @bhuvip9204 Рік тому

      Ok but my question is to find SD u have to do repeatability. So while performing repeatability how many gm weight should I use? If i have 220gm balance..than how will I know what weight should i use for repeatability to get SD. Please do reply. Thanks

  • @prasannakumar1614
    @prasannakumar1614 9 місяців тому

    Why 2000 , why not 1000 sir

  • @bestisyettocome2645
    @bestisyettocome2645 Рік тому

    Sir what is the difference between smallest net weight and minimum operating weight ??