I think of frequency as being synonymous with rate, and amplitude with delta P, however I do understand what you’re driving at if the Hertz (frequency) are fixed. The goal being to increase ventilation.
Thank you for such great simplified presentation I need to make sure from you... Amplitude means respiratory rate or not? Please keep educating us with your simple way.
Frequency - This determines the frequency of the pressure swings and usually varies from 7-15 Hz (1 Hz = 60 cycles/min) in neonatal practice. Ventilation is most efficient if the frequency matches the natural resonant frequency of the diseased lungs. It is a very powerful tool for CO2 elimination and should only be altered after careful consideration and discussion with senior experienced personnel. Usual magnitude of change 1 Hz / step
I think of frequency as being synonymous with rate, and amplitude with delta P, however I do understand what you’re driving at if the Hertz (frequency) are fixed. The goal being to increase ventilation.
Thank You for a great explaination. Please keep making lecture sir....
Excellent presentation
need more videos on neonatology
Well done Mohammed thank you x
We need this presentation pdf
Hi
What is start of Amplitude
Thank you for such great simplified presentation
I need to make sure from you... Amplitude means respiratory rate or not?
Please keep educating us with your simple way.
Amplitude is your delta creating Vt. Hz is your frequency that’s why it’s 60 breaths per 1 Hz
Thank you
جزاك الله خيراً
Thanks alot. What's the initial setting for amplitude. Is it double MAP as many do?
Hi
You should be guided by good chest wall movement ( wobbles) as well as good chest expansion in CXR.
Excellent
Thanks a lot 🌹
Amplitute means change in pressure not the respiratory rate I think?
You are correct, amplitude is delta P.
Great 👍👍👍
why not to change the frequency?
Frequency - This determines the frequency of the pressure swings and usually varies from 7-15 Hz (1 Hz = 60 cycles/min) in neonatal practice. Ventilation is most efficient if the frequency matches the natural resonant frequency of the diseased lungs.
It is a very powerful tool for CO2 elimination and should only be altered after careful consideration and discussion with senior experienced personnel. Usual magnitude of change 1 Hz / step
Crisp