II How to oxygen (O2) analyzer calibration II Zirconia cell replaced II
Вставка
- Опубліковано 16 вер 2024
- In this vedio I am sharing the Yokogawa Integral type oxygen analyzer Maintainance and calibration.
Make- Yokogawa Integral Type O2 analyzer
Model - ZR202G
II replaced Zirconia cell procedure . carefully removal and installation II
@@Calibration@@
II zero calibration by standard gas (0.4% o2)II
II span calibration by open port in atmosphere (21%) II
@@analyzer function code @@
For check zero & span value - b01 & b02 respectively.
@@for enter in calibration mode function code is b10@@
II firstly span calibration then zero calibration.
calibration completed II
so good keep it up...
Good work keep it up..
Best vedio for calibration
Thanks
Good job vire
Good job bro
Thanks for your comment 🙏🏻🙏🏻🙏🏻
👆☝️👍🖐🤘good morning sir/very good job👍👌✌🖖🖕👐👏👋
Thanks for comment 🙏🏻
thanks so much ❤
Thanks 🙏🏻
Bahut help full video hai ❤❤❤
But sir ji standard gas kon si hoti hai or kitne % main aati hai
bhut achhe se samjhaya hindi m ydi sir standard value dene pr .4 dene pr . 2 ya .3 aay to kya krna chahiy
Aapne vedio me dekha hoga jab maine zero calibration kiya to standard value se match nahi ho raha tha isiliye to calibration kiya maine . Calibration me jo error/ drift hota hai usko adjust kar leta hai analyzer.
Any other doubt then ask me
ok sir automatic arjest ho jata h . sir vibrationn sensor key phasor ko v samjhaiye
sir conductivity analyzer k bare m v banaiye
Sure Jaroor bataunga vedio about conductivity
Fantastic
Bahut shai Boss
Thanks bhai
Please bataia ye analyzer kha pe installed h mtlb kon se unit k exact kis location pe....
Ok ✅ thanks for your comment and suggestions
Nice vedio
👍👍
hii good video.
can you please tell me why you given first span gas air and then 0.4 zero gas?
Thanks for your comments 🙏🏻 it is not necessary you can do any first
Good
Sir no2 gas zero ke liye span ke liye reference gas kya use karte hai
Sir span cal 8.0 gas mai hota hai na,plz reply
Thoda theory part bhi cover kijiye Boss
Sure sure
Thank you for your helpful video and I have a confused question. When you remove the detector, the boiler is shutting down or still operating? Thank you so much!
Thankyou so much for supporting my channel and like videos . Boiler in Operating Condition . If any other question then ask me
@@PowerPlantInstrumentation thanks for your response! The boiler is operating, the pressure of flue gas inside the boiler isn't influenced to boiler ? I mean it is maybe exploded or another problem occurs
@@InstrumentationTechnicianVP No, there is negative pressure in the boiler. So there is no problem in getting the sensor out.
@@PowerPlantInstrumentation thank you so much
🙏🏻
Thank you for sharing the video! When you set the atmosphere on the open port to 21% O2, is the plant at sea level? If not, can this negatively affect the process the instrument is in?
Yes atmosphere oxygen concentration at seal level .
Calibration successfully both span and zero but in load time value not correctly when load increase or decrease.plz suggest
check Zirconia Cell
Sir SOx and NOx calibration ka bhi vedio banona (Forbes Marshall)
Ok
Which port should be open for span calibration and which port should I give reference air 1-1.5kg/cm2. Pl explain clearly
I have shown in this vedio -only CAL port used for zero & span calibration.
Zirconia cell cost?
Where did this analyser take place in this video ?
Air Pre Heater Inlet (APH) of Power plant
Services engineer no
How much supply given to heater
230vac
Unit kya hi ?
Percentage of oxygen concentration. (%)
How much suppl
230vac supply given to heater sir