Це відео не доступне.
Перепрошуємо.

Error Code 551 - Alcon Infiniti

Поділитися
Вставка
  • Опубліковано 30 лип 2023
  • Error Code 551 on the Alcon Infiniti refers to the AquaLaz system that was designed for the AquaLaz Handpiece that used warm BSS pulse to breakup the cataract. So if you get Error Code 551 that means your AquaLaz is still activated and you will need to de-activate that system.
    To learn more about Error Code 551, see video below!
    See the full Error Code Playlist for more Error Codes that may come up during service of Alcon Centurion - • Error Codes
    CONTACT US:
    6890 Hawthorn Park Drive
    Indianapolis, IN 46220
    Phone:
    317.663.3158
    855.233.4050
    Email:
    sales@medisurg.com
    info@medisurg.com
    www.MEDiSURG.com

КОМЕНТАРІ • 3

  • @abdulmuhaimenhamuda6016
    @abdulmuhaimenhamuda6016 Рік тому +1

    Dear sir
    Could you help me with my Alcon Infiniti it gave 135 error message
    Fluidic not available, please restart the system ….
    Although the machine has not been used for long time
    Can you help me with that ?

    • @anasfuad6575
      @anasfuad6575 Рік тому

      This error code 135 means that the spacer in the fluidics module needs to be replaced.

    • @MEDISURGvid
      @MEDISURGvid  Рік тому +1

      The vent valve is stuck. You can open the left side of the system and get access to the solonoid valves. One of them are stuck. You may need to replace the rubber spacers to ensure it will not go to the error state.. During startup, the solonoid valves are activated and if they are stuck, it will give this error.